Validation of Aptamer-Protein Interactions by Biolayer Interferometry: A Study of 88 Candidate Molecules.
Elizaveta S Ershova, Natalia V Pozdnyakova, Aleksandra Yu Egorshina, Elena V Yakushenko, Igor V Dobrokhotov, Nadezhda P Lyamina, Bella B Gagulaeva, Alexey Moskalev
Petrovsky National Research Center of Surgery, Institute of Biology of Ageing and Healthy Longevity Medicine with Clinic of Predictive Medicine, 119435 Moscow, Russia. · SistemaBioTech, LLC, 1-ya Kuryanovskaya Str. 34, 109235 Moscow, Russia.
Key points
From the paper's conclusion- Local fitting returns internally consistent KD for data incompatible with a 1:1 model; we propose isotherm verification and antibody validation of every preparation as minimal requirements.
Abstract
An aptamer's low mass confines the label-free response to hundredths of a nanometer, where artifacts mimic binding yet yield computable dissociation constants. To define and test a priori criteria distinguishing binding from artifact, 88 computationally designed aptamers against seven human proteins were studied by biolayer interferometry in two orthogonal configurations, with control antibodies. Six criteria, fixed in advance, are reported as a priori decision rules, not validated requirements. Control antibodies validated 6 of 8 preparations; for the other 68 pairs, the maximum response was 0.019-0.219 nm, bounding KD at ≥8.6 µM, 120-fold above the 70 nM threshold. Two of three signals clearing it gave exponents far above the 1:1 expectation (1.04 vs. 0.31; 0.85 vs. 0.41); the third failed parameter consistency (KD scatter ×28). Rankings on two PCSK9 preparations did not correlate (ρ = -0.26); predicted and experimental KD differed by 3-7 orders of magnitude (ρ = 0.497, n = 25). No aptamer reached the predefined affinity under the conditions tested. Local fitting returns internally consistent KD for data incompatible with a 1:1 model; we propose isotherm verification and antibody validation of every preparation as minimal requirements. No positive-control aptamer was available, limiting interpretation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Widespread omission of aptamer-target binding verification in aptasensor development: Consequences for sensor performance and the need for a Kd gateBiosensors and Bioelectronics · 2026
- A Practical Guide of Choosing and Validating Aptamers for Biosensors.ACS sensors · 2026
- Beyond Sensitivity: Materials and Biointerface Design Rules for Aptamer-Enabled Smart Contact Lenses.ACS materials Au · 2026
- Teaching Bio‐Layer Interferometry: Protein–Protein Versus Protein–Small Molecule Interaction Analysis on an Octet R8Biochemistry and Molecular Biology Education · 2026
- Retention-Based, Dissociation-Sensitive Pre-Screening of Aptamers Using a Centrifugal Microspin Filter.Biosensors · 2026
- Systematic Evolution of Ligands by Exponential Enrichment to Sensor: Translating an Aptamer Into a Functional Electrochemical SensorAnalysis & Sensing · 2026