Calibration Strategies for Correcting Matrix Effects in ICP-MS Multi-Element Analysis Using a Certified Reference Material Followed by Application to Brown Crab (Cancer Pagurus) Tissues.
Dongli Dong, Raquel Cama-Moncunill, David van Acken, Selene Pedrós-Garrido, Daniel Hurley, James G Lyng
UCD School of Agriculture and Food Science, University College Dublin, D04 V1W8 Dublin, Ireland. · UCD School of Earth Sciences, University College Dublin, D04 V1W8 Dublin, Ireland.
Key points
From the paper's conclusion- This study evaluated five calibration strategies for the determination of As, Cd, Cr, Ni, Pb and Se in seafood matrices using a certified reference material (CRM, ERM®-CE278k, mussel tissue) and brown crab (Cancer pagurus) hepatopancreas and claw meat.
Abstract
Accurate multi-element quantification in seafood by inductively coupled plasma mass spectrometry (ICP-MS) can be compromised by matrix effects, particularly when aqueous-based calibration is applied to complex biological digests. This study evaluated five calibration strategies for the determination of As, Cd, Cr, Ni, Pb and Se in seafood matrices using a certified reference material (CRM, ERM®-CE278k, mussel tissue) and brown crab (Cancer pagurus) hepatopancreas and claw meat. External calibration (EC), EC with internal standard (EC-IS), standard addition (SA), SA with internal standard (SA-IS), and single-level SA-IS (SLSA-IS) were first compared using the CRM. Matrix effects were quantified from the slope ratio between SA and EC calibration curves. Significant signal suppression of -27.75% was observed for Se with the 95% confidence interval (CI) from -41.2% to -14.3% and -18.1% for Cd (95% CI: -25.2% to -11.0%), while As showed a borderline suppression trend (-15.1 ± 6.2%, p = 0.052) with the 95% CI from -30.6 to 0.3%. External calibration and EC-IS gave acceptable precision but underestimated As, Cd and Se, with biases of -25.66%, -22.63% and -35.54% for EC and -20.16%, -17.03% and -31.47% for EC-IS, respectively. Standard addition-based calibration improved trueness, and SA-IS produced the closest agreement with the certified elemental concentration, with biases of -2.91%, 1.20%, -10.34%, 2.92%, -0.30% and -3.42% for As, Cd, Cr, Ni, Pb and Se, respectively. The performance of SLSA-IS was spike-dependent (1-50 µg/L), with higher spike levels yielding the most accurate predictions. Subsequently, SA-IS was explored in hepatopancreas and claw meat from two brown crabs, providing satisfactory determinations for Cd, Cr, Ni, Pb and Se, whereas As quantification was limited by high native concentrations exceeding the standard-addition range in the present study. Cross-matrix evaluation showed that agreement between MMC and SA-IS was element-, matrix- and batch-dependent, with crab tissue-based MMC generally showing closer agreement than CRM-based MMC. Therefore, MMC deserves further investigation for selected applications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Quantitative Detection of Toxic Elements in Food Samples by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)Processes · 2025
- A Greener Acid-free Microwave-Assisted Digestion Method for Multielement Analysis of Fish Tissues by ICP-MS.ACS omega · 2026
- Evaluation of extraction capabilities of protease, proteinase K, and TMAH for nanoparticle characterization in biological CRMs by spICP-MS.Analytical and bioanalytical chemistry · 2026
- A frozen-suspension external calibration strategy for quantitative imaging of Ce in fresh animal soft tissues by CLA-ICP-MS.Talanta · 2026
- EXPRESS: ICP-MS Analysis of Pipe Tobacco Using Calibration by Proxy.Applied spectroscopy · 2026
- Bioaccumulation and exposure to potentially toxic elements (PTEs) in marine species commercialized for human consumption in the Southern Region of Rio de Janeiro (SE, Brazil)Marine Pollution Bulletin · 2026