Evaluation of extraction capabilities of protease, proteinase K, and TMAH for nanoparticle characterization in biological CRMs by spICP-MS.
Filip Gregar, Doaa Abouelenein, Katrin Loeschner, Zuzana Gajdosechova
Metrology Research Center, National Research Council Canada, 1200 Montreal Road, Ottawa, ON, K1A 0R6, Canada. · Research Group for Analytical Food Chemistry, National Food Institute, Technical University of Denmark, 2800 Kgs., Lyngby, Denmark.
Key points
From the paper's conclusion- Overall, this work provides a systematic comparison of commonly used extraction strategies for the analysis of naturally occurring NPs in biological CRMs, identifies current limitations associated with sample preparation, and provides guidance for future method development and the design of biological CRMs for NP characterization.
Abstract
Reliable method development and validation rely on certified reference materials (CRMs) that ideally provide accurate analyte values independent of sample preparation and analytical methodology. Although inorganic nanoparticles (NPs) are increasingly being investigated in biological systems, biological matrix CRMs certified for NP characterization are not yet available. In this study, five biological CRMs (DOLT-4, DOLT-5, DORM-4, DORM-5, and TORT-3) were extracted using three commonly employed protocols based on proteinase K, protease, and tetramethylammonium hydroxide (TMAH). The resulting extracts were analyzed by single-particle ICP-MS (spICP-MS) to determine particle number concentrations (PNCs) and particle size distributions (PSDs) of Ag-, Al-, Cu-, Cr-, Fe-, Hg-, Se-, Si-, Ti-, and Zn-containing nanoparticles. Both PNCs and PSDs varied substantially among the extraction protocols, demonstrating that sample preparation is a major source of variability in NP characterization. In addition to differences in extraction performance, elevated NP backgrounds were observed for several elements in the TMAH reagent, highlighting the importance of evaluating reagent-derived contamination when interpreting spICP-MS results. Investigation of Hg- and Se-containing NPs in DOLT-4, DORM-5, and TORT-3 further demonstrated that incomplete tissue digestion and residual matrix fragments complicate particle identification and interpretation. Overall, this work provides a systematic comparison of commonly used extraction strategies for the analysis of naturally occurring NPs in biological CRMs, identifies current limitations associated with sample preparation, and provides guidance for future method development and the design of biological CRMs for NP characterization.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of reagents for suspension-assisted total reflection X-ray fluorescence element analysis in biomedical samplesAnalytical and Bioanalytical Chemistry · 2026
- Preparation and certification of novel high-protein food certified reference materials.Analytical and bioanalytical chemistry · 2026
- Application of enzymatic hydrolysis for the isolation of titanium dioxide nanoparticles from human hair for sp-ICP-MS analysis.Analytical and bioanalytical chemistry · 2026
- Physicochemical characterization of nanoparticles in highly diluted preparations and exploratory plasma proteomic correlates in an N-of-1 study.Nanomedicine : nanotechnology, biology, and medicine · 2026
- Development of a Flow-through Cell for Ultrasonic Extraction (UE)─Single Particle (SP)─ICP-MS─an Approach for Nano/Microparticle Elemental and Isotopic Analysis.Analytical chemistry · 2026
- Calibration Strategies for Correcting Matrix Effects in ICP-MS Multi-Element Analysis Using a Certified Reference Material Followed by Application to Brown Crab (Cancer Pagurus) Tissues.Foods (Basel, Switzerland) · 2026