Key points
From the paper's conclusion- Conclusions: Under the in vitro conditions tested, the prosthetic screw incorporating a 1.0 mm secondary retrieval head demonstrated reliable retrieval performance, supporting its potential as a simple retrieval system.
Abstract
Background/Objectives: This study aimed to present the design of a novel abutment screw head featuring an integrated retrieval mechanism and to evaluate the retrievability of four prototype designs. Methods: In this in vitro mechanical study, the modified screws were produced by replicating the macrogeometry of the Brånemark and Klockner VEGA abutment screws and incorporating a secondary screw chamber below the original one, with a 1.0 mm or 0.9 mm hexagonal design. Sixty screws were manufactured, fifteen for each group: BC0.9, BC1.0, KC0.9, and KC1.0. Retrieval success and the relationship between peak tightening and loosening torque were evaluated (α = 0.05). Results: All screws tolerated tightening at a torque 20% above the manufacturer's recommendation. Most specimens from BC0.9 and KC0.9 failed during loosening, while all specimens in BC1.0 and KC1.0 were successfully retrieved. Regarding peak tightening torque and the loosening-to-tightening ratio, neither prototype exhibited a statistically significant correlation. BC1.0 demonstrated a statistically significant positive correlation between peak tightening torque and peak loosening torque, whereas KC1.0 showed a weaker relationship; detailed correlation coefficients and p-values are reported in the Results. Conclusions: Under the in vitro conditions tested, the prosthetic screw incorporating a 1.0 mm secondary retrieval head demonstrated reliable retrieval performance, supporting its potential as a simple retrieval system. These findings represent a proof-of-concept and require further mechanical and clinical investigation before clinical application.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Stepwise Technique for Retrieval of a Seized and Stripped Dental Implant Cover Screw: A Case Report.Reports (MDPI) · 2026
- Effect of Tightening Protocol on Reverse Torque and Plastic Deformation of Dynamic and Conventional Titanium Abutment Screws.Bioengineering (Basel, Switzerland) · 2026
- Titanium Base-Free Multi-Unit Abutment Connections: A Critical Review of Prosthetic Screw Design, Biomechanical Behavior, and Clinical PerformanceMaterials · 2026
- A Conservative Chairside Technique for Retrieval of a Fractured Implant Abutment Screw Using a Modified Hand Instrument: A Case ReportOral Surgery · 2026
- Influence of different abutment materials on torque maintenance and screw loosening in implant-supported prostheses.Bioinformation · 2026
- A comparative analysis of efficacy of torque limiting devices provided by different dental implant manufacturersBMC Oral Health · 2026