A New Protocol for Evaluating Maximum Respiratory Pressures in Individuals With Multiple Sclerosis: Concurrent Validity and Test-Retest Reliability.
Raúl Fabero-Garrido, Sandra Aguilar-Zafra, Tamara Del Corral, Ibai López-de-Uralde-Villanueva
Department of Physiotherapy, Faculty of Nursing, Physiotherapy and Podiatry, Complutense University of Madrid, Madrid, Spain (R.F.-G.); Departamento de Fisioterapia, Facultad de Ciencias de la Salud, Motion in Brains Research Group, Centro Superior de Estudios Universitarios La Salle, Universidad Autónoma de Madrid, Spain (S.A.-Z.); Téxum S.L. Physiotherapy Center, Coslada, Spain (S.A.-Z.); and Department of Physiotherapy, Faculty of Nursing, Physiotherapy and Podiatry, Universidad Complutense de Madrid; IdISSC, Madrid, Spain (T.d.-C. and I.L.d.U.V.).
Key points
From the paper's conclusion- The concurrent validity of the 1RM-based protocol was good to excellent with respect to the reference test (MIP: day 1, n = 49, intraclass correlation coefficient [ICC (95% confidence intervals)] = 0.87 (0.78-0.93); day 2, n = 48, ICC = 0.91 (0.82-0.95); MEP: day 1, n = 49, ICC = 0.82 (0.60-0.91); day 2, n = 49, ICC = 0.88 (0.64-0.95)). The test-retest reliability was good for MIP [n = 46, ICC = 0.88 (0.79-0.93)] and excellent for MEP [n = 48, ICC = 0.90 (0.83-0.94)]. The MDC(90-95) for MIP ranged from 23.7 to 28.2 cmH2O, and for MEP ranged from 19 to 22.5 cmH2O.
Abstract
BACKGROUND AND PURPOSE: There is a need for new and inexpensive methods for the maximum inspiratory and expiratory pressures (MIP/MEP) test. The one-repetition maximum (1RM) test is an easy and cost-effective method to measure maximal strength. The purpose of this study was to validate a MIP/MEP test protocol based on the 1RM test, its test-retest reliability, and minimal detectable change (MDC) in individuals with multiple sclerosis. METHODS: Fifty-two individuals (53.6 ± 10 years; 5.7 ± 1.8 points on Expanded Disability Status Scale) (46% relapsing-remitting; 23% primary progressive; 31% secondary progressive) attended two appointments 7 to 10 days apart. MIP/MEP tests were performed using a threshold valve device (1RM-based protocol) and a digital manometer (reference test). The 1RM-based protocol included an incremental phase (load increase [30 cmH2O] to achieve task failure) and a halving phase (load halfway between the lowest failed attempt and the last valid attempt was prescribed). RESULTS: The concurrent validity of the 1RM-based protocol was good to excellent with respect to the reference test (MIP: day 1, n = 49, intraclass correlation coefficient [ICC (95% confidence intervals)] = 0.87 (0.78-0.93); day 2, n = 48, ICC = 0.91 (0.82-0.95); MEP: day 1, n = 49, ICC = 0.82 (0.60-0.91); day 2, n = 49, ICC = 0.88 (0.64-0.95)). The test-retest reliability was good for MIP [n = 46, ICC = 0.88 (0.79-0.93)] and excellent for MEP [n = 48, ICC = 0.90 (0.83-0.94)]. The MDC(90-95) for MIP ranged from 23.7 to 28.2 cmH2O, and for MEP ranged from 19 to 22.5 cmH2O. DISCUSSION AND CONCLUSION: This study validated a new 1RM-based protocol for the MIP/MEP tests using a respiratory muscle training threshold valve device in individuals with multiple sclerosis, showing good to excellent concurrent validity compared with the reference test, as well as good to excellent test-retest reliability. The MDC reported can be interpreted and applied in the clinical setting.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Bench Agreement of Two Units of a Low-Cost Digital Manometers with a Clinical Respiratory Pressure Comparator: A Laboratory-Based Comparison.Diagnostics (Basel, Switzerland) · 2026
- Feasibility, Reliability, and Convergent Validity of the Five Times Sit-to-Stand Test via Telehealth in Multiple Sclerosis.International journal of MS care · 2026
- Construct validity and intrasession intra-rater reliability of the stair and step tests in individuals with multiple sclerosis.Physiotherapy theory and practice · 2026
- Individualized respiratory exercise device originated in Türkiye: design, prototype, and first resultsTURKISH JOURNAL OF MEDICAL SCIENCES · 2026
- Test–Retest Reliability and Convergent Validity of the 6-Minute Pegboard and Ring Test in Obese IndividualsMeasurement in Physical Education and Exercise Science · 2025
- A multi-dermatome computer-controlled electrical perceptual threshold (multi-dermatome CCEPT) test: reliability assessment and development of a preliminary healthy dermatomal reference profileBioMedical Engineering OnLine · 2026