Automated Versus Manual Occlusion Pressure at 100 ms With Hamilton C6 Using the Software Version 1.2.3: From Bedside to the Bench and Back to Bedside.
Flora Delamaire, Jérémy Caldonazzo, Arnaud Gacouin, Rémi Coudroy, Nicolas Terzi
Médecine Intensive - Réanimation, CHU de Rennes, Rennes, France. · Médecine Intensive Réanimation, CHU de Poitiers, Poitiers, France.
Key points
From the paper's conclusion- Automatic P0.1 estimation by the Hamilton C6 (software version 1.2.3) is inaccurate during SBTs. Manual occlusion remains essential for reliable P0.1 assessment at the bedside.
Abstract
OBJECTIVES: Occlusion pressure at 100 ms after the start of inspiratory effort (P0.1) is a marker of respiratory drive and a predictor of weaning outcomes in mechanically ventilated patients. Some ventilators, like the Hamilton C6 (Hamilton Medical AG, Bonaduz, Switzerland), estimate P0.1 automatically (P0.1vent) without performing an actual airway occlusion, potentially leading to inaccuracies. SETTING: This prospective clinical and bench study aimed to assess the accuracy of P0.1vent compared with the manual occlusion method (P0.1occ) during spontaneous breathing trials (SBT) and following a reventilation period with varying ventilator settings. DESIGN AND PATIENTS: The clinical part of the study included patients undergoing SBT on Hamilton C6 (software version 1.2.3) with zero pressure support (PS) and zero end-expiratory pressure (PEEP) and reventilation (PS +8 cm H2O, PEEP +5 cm H2O). P0.1vent and P0.1occ were collected every 15 minutes. Bench part evaluated the P0.1vent and P0.1occ of the Hamilton C6 (software version 1.2.3) connected to a lung simulator (ASL 5000) creating different breathing efforts across different lung mechanics and varying PS and PEEP levels. MEASUREMENT AND MAIN RESULTS: In patients, P0.1vent values were significantly higher than P0.1occ (median, 11.0 vs. 1.0 cm H2O; p < 0.001), with a mean bias of 8.5 cm H2O. Similar overestimations were observed during the reventilation period (6.0 vs. 1.0 cm H2O; mean difference, 6.5 cm H2O; p < 0.001). Bench results confirmed the variable accuracy of P0.1vent found in the clinical part of the study, with biases ranging from nearly perfect (0.2 cm H2O) under high PS to 2.3 cm H2O under low PS. CONCLUSIONS: Automatic P0.1 estimation by the Hamilton C6 (software version 1.2.3) is inaccurate during SBTs. Manual occlusion remains essential for reliable P0.1 assessment at the bedside.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Assessment of Respiratory Mechanics and Inspiratory Effort During Spontaneous Breathing Trials to Predict Extubation Failure in High-Risk PatientsAmerican Journal of Respiratory and Critical Care Medicine · 2025
- Monitoring of breathing effort through pressure time product measurement using airway occlusion pressure (PTPPOCC study).BMC anesthesiology · 2026
- Tidal swing of endotracheal tube cuff pressure as a measurement of inspiratory effort and lung stress during pressure support ventilation: a proof-of-concept study.Anesthesiology · 2026
- Identification and Characterization of Reverse Triggering in Acute Hypoxemic Respiratory Failure Using an Automated ToolRespiratory Care · 2026
- Association of Positive Airway Pressure With Non-Smoke Spontaneous Individual Contrast During Simulated Valsalva LoadingEuropean Heart Journal - Imaging Methods and Practice · 2026
- The MIME framework for non-invasive longitudinal assessment of inspiratory effort during pressure support ventilation: a physiological validation study.Critical care (London, England) · 2026