Ultra-High-Resolution Imaging of Rectal Cancer with Photon-Counting Detector Computed Tomography: An Exploratory Pilot Study on the Impact of Slice Thickness and Monoenergetic Reconstruction Level on Image Quality.
Alexander Herold, Andreas Strassl, Daniela Beitzke, Philipp Stelzer, Christian Wassipaul, Theresa Dachs, Bence Nemeth, Florian Lindenlaub,
Department of Biomedical Imaging and Image-Guided Therapy, Medical University of Vienna, Waehringer Guertel 18-20, 1090 Vienna, Austria. · Division of Visceral Surgery, Department of General Surgery, Comprehensive Cancer Center Vienna, Medical University of Vienna, Waehringer Guertel 18-20, 1090 Vienna, Austria.
Key points
From the paper's conclusion- Overall quality was rated favorably in 82.9% at 70 keV versus 45.6% at 45 keV.
Abstract
Background/Objectives: To evaluate, in an exploratory pilot study, the influence of slice thickness and monoenergetic (ME) level on quantitative and qualitative image quality of ultra-high-resolution (UHR) photon-counting detector CT (PCD-CT) in rectal cancer (RC). Methods: This prospective pilot study enrolled 21 patients with suspected RC undergoing pre-treatment PCD-CT between July 2023 and February 2025. All underwent a QuantumPlus portal venous phase, whose 3 mm/60 keV reconstruction served as the standard reference, plus a separate pelvic UHR acquisition. Sixteen UHR reconstructions were generated (0.6-3 mm; 45-70 keV). Signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and noise were analyzed using linear mixed-effects models. Three readers independently rated overall image quality, perceived noise, mesorectal fascia (MRF) sharpness and lesion conspicuity relative to the reference; ratings were analyzed using mixed model ordinal regression. The primary analysis was the within-UHR comparison of reconstruction settings; the comparison with the reference, which differed in contrast phase and acquisition parameters, was descriptive. Results: Nineteen patients (mean age 61.7 ± 10.7 years, 73.7% male) were included. CNR showed a significant slice thickness-ME level interaction (p < 0.001), ranging from 4.85 ± 2.43 (3 mm/45 keV) to 0.86 ± 1.04 (0.6 mm/70 keV). SNR increased and noise decreased with thicker slices, while both decreased with higher ME levels (all p < 0.001). ME level influenced all four qualitative parameters (all p < 0.001); slice thickness significantly affected overall quality and perceived noise (p < 0.001) but not MRF sharpness or lesion conspicuity (p ≥ 0.115). Overall quality was rated favorably in 82.9% at 70 keV versus 45.6% at 45 keV. Lesion conspicuity reached 71.9-86.0% favorable ratings, whereas perceived noise was consistently rated worse than the reference, and UHR SNR remained below the reference in all settings. In a descriptive pooled summary, 3 mm/60 keV (68.0%) and 2 mm/70 keV (66.2%) were rated favorably most often, with overlapping confidence intervals. Conclusions: Slice thickness and ME level markedly influence UHR PCD-CT image quality in RC. One candidate working reconstruction for future validation is 3 mm at 60 keV, with 2 mm at 70 keV as a plausible alternative. Because the reference differed in acquisition and contrast phase and no diagnostic endpoint was assessed, these results do not establish a diagnostic benefit of UHR PCD-CT.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Detection of hypovascular colorectal liver metastases with photon-counting detector CT versus energy-integrating detector CT: a blinded intra-patient comparison of routine clinical protocols using an MRI-anchored reference standardFrontiers in Oncology · 2026
- Photon-Counting CT With Low-keV Virtual Monoenergetic Images for Detecting Muscle Invasion in Bladder Cancer: A Prospective Study.AJR. American journal of roentgenology · 2026
- Image quality and dose performance of photon-counting CT in multiple myeloma: a comparative clinical study.Journal of cancer research and clinical oncology · 2026
- Small-pixel Acquisition Unmasks "Bucket Effects" of Iterative Reconstruction in Ultra-High-Resolution Photon-Counting CT.Academic radiology · 2026
- Image quality and artefact reduction for tantalum clips in ocular proton therapy using photon-counting CT.European radiology experimental · 2026
- Optimization of reconstruction strategies for head and neck computed tomography angiography using photon-counting computed tomography: A comparative study with energy-integrating detector computed tomography.European journal of radiology · 2026