Image quality and dose performance of photon-counting CT in multiple myeloma: a comparative clinical study.

Matthias Michael Woeltjen, Samir Abu Salim, Denise Schoenbeck, Saher Saeed, Nina Pauline Haag, Julius Henning Niehoff, Jan Robert Kroeger, Jan Borggrefe,

Department of Radiology, Neuroradiology and Nuclear Medicine, Johannes Wesling University Hospital, Ruhr University Bochum, Bochum, Germany. Matthiasmichael.Woeltjen@muehlenkreiskliniken.de. · Department of Radiology, Neuroradiology and Nuclear Medicine, Johannes Wesling University Hospital, Ruhr University Bochum, Bochum, Germany.

Observational · n=47

Key points

From the paper's conclusion
  • In non-contrast whole-body imaging of patients with multiple myeloma, the evaluated PCCT system provided superior image quality and significantly lower radiation-dose indices than the historical EID-CT system. These findings reflect the combined performance of the imaging systems and their respective acquisition and reconstruction settings.

Abstract

PURPOSE: To compare image quality and radiation-dose indices between photon-counting computed tomography (PCCT) and conventional energy-integrating detector CT (EID-CT) examinations in patients with multiple myeloma. METHODS: This retrospective intraindividual study included 47 patients with non-contrast whole-body PCCT examinations (NAEOTOM Alpha, Siemens Healthineers) and prior EID-CT examinations (SOMATOM Sensation 64, Siemens Healthineers). Examinations were performed using the respective clinical protocols, which differed in acquisition and reconstruction settings. Quantitative image analysis included region-of-interest measurements in the lungs, spleen, muscle, and adipose tissue. Image noise and signal-to-noise-ratio (SNR) were calculated. For qualitative assessment, two radiologists independently evaluated randomized and anonymized PCCT and EID-CT scans of the same patients using a 5-point Likert scale to rate overall image quality, sharpness, noise, and osteolytic lesion conspicuity. Qualitative ratings were compared separately for each reader using two-sided paired Wilcoxon signed-rank tests. Differences in CT-dose-index, dose-length-product, and quantitative image metrics were analyzed using two-sided paired t-tests. RESULTS: Quantitative analysis demonstrated significantly lower image noise in PCCT compared to EID-CT in the spleen, muscle, and adipose tissue (p < 0.0001), but not in lung tissue (p = 0.27). SNR was significantly higher with PCCT in all tissues except the lungs (p < 0.0001; lungs: p = 0.22; adipose tissue: p = 0.02). Qualitative assessment showed that PCCT was rated significantly better than EID-CT for overall image quality, image sharpness, noise, and osteolysis conspicuity (p < 0.0001 for all categories). Radiation dose parameters (CTDIvol and DLP) were significantly lower with PCCT than with EID-CT (p < 0.0001). CONCLUSIONS: In non-contrast whole-body imaging of patients with multiple myeloma, the evaluated PCCT system provided superior image quality and significantly lower radiation-dose indices than the historical EID-CT system. These findings reflect the combined performance of the imaging systems and their respective acquisition and reconstruction settings.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers