Multidimensional Tracing of Error Sources and Structured Corrective Strategies for Leukocyte Count and Differential on Automated Hematology Analyzers.
Ping Deng, Lianhui Yu, Haoting Zhan, Lin Zheng, Yipei Jing, Jiaying Fang, Min Zhang, Wei Wu
Department of Laboratory Medicine, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences & Peking Union Medical College, Beijing 100730, China.
Key points
From the paper's conclusion- We categorized multidimensional error sources into specimen-related interferents, leukocyte morphological and developmental abnormalities, and preanalytical or systematic factors, and established a practical diagnostic workflow spanning from anomaly recognition to result reporting.
Abstract
Leukocyte count and differential are core indicators for clinical infection assessment, hematological disease diagnosis, and therapeutic monitoring. Currently, clinical testing heavily relies on automated hematology analyzers. Influenced by factors such as specimen characteristics, limited analyzer recognition capability, and improper manual operations, automated testing often produces unreliable results, interfering with clinical decision-making and threatening patient safety. Existing studies remain fragmented, lacking a multidimensional error-tracing framework and practical corrective strategies covering the entire testing workflow. A narrative review was conducted to summarize the causes, identification methods, and corrective strategies for unreliable leukocyte results. Relevant literature up to August 2026 was searched in PubMed/MEDLINE and CNKI. We categorized multidimensional error sources into specimen-related interferents, leukocyte morphological and developmental abnormalities, and preanalytical or systematic factors, and established a practical diagnostic workflow spanning from anomaly recognition to result reporting. By incorporating both legacy and current analytical systems, we compared the characteristic scattergram patterns of various interferents across analyzers with different detection principles, broadening the applicability to diverse clinical laboratories. To better ensure the reliability of corrective strategies and provide diverse alternative options, we not only summarized widely accepted routine practices but also included case-based experimental methods. The proposed multidimensional error-tracing system and practical diagnostic workflow significantly improve analytical accuracy and reduce clinical risks. However, current hematology analyzers cannot automatically identify or correct complex interferences. Future systems must integrate multi-platform advantages, optimize complex cell recognition capabilities, and enrich multidimensional scattergram information to overcome existing analytical limitations. Furthermore, integrating artificial intelligence technology holds immense potential to drive automated hematology analysis toward a more intelligent and precise direction.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Preanalytical interventions to reduce hemolysis and improve specimen quality: a systematic review and meta-analysis.Laboratory medicine · 2026
- Real-World Evaluation of the Sysmex XN-9100 WDF-WPC Workflow: Diagnostic Performance, False-Negative Burden, and Potential Optimization of Smear Review Workload.Diagnostics (Basel, Switzerland) · 2026
- Utility of WBC Scatterplots and Suspect Flags Generated by Beckman Coulter LH-750 Hematology Analyzer in the Characterization of Leukemias and Related Hematological Malignancies.Indian journal of hematology & blood transfusion : an official journal of Indian Society of Hematology and Blood Transfusion · 2026
- Why do hemolysis interference studies produce different results? A proposal for standardized experimental design and reportingClinical Chemistry and Laboratory Medicine (CCLM) · 2026
- Beyond Analytical Agreement: Integrating Analytical Performance and Algorithm-Assisted Morphological Interpretation in a Comparative Evaluation of the Mindray BC-7800/BC-7900 and Sysmex XN-9100 Hematology Platforms.Diagnostics (Basel, Switzerland) · 2026
- Reduction in Sample Recollection Number Following the Implementation of Automated Pre-Analytical Interference Detection in Coagulation Testing.Annals of clinical biochemistry · 2026