Impact of patient centering on radiation dose and image noise in computed tomography: Noise reduction potential of adaptive statistical iterative reconstruction
Yassine Oustous, Aziz Oustous, Zouhair Airouss, M. El Katib, F. Chehab, N. Ech-cherif El Kettani, M. Jiddane
Key points
From the paper's conclusion- Using fixed scanner settings, we measured how such vertical offsets change image noise and the dose at given points, and if the extra noise can be compensated by Adaptive Statistical Iterative Reconstruction (ASIR), which offsets the extra noise.
Abstract
In computed tomography (CT), a patient placed above or below the isocenter is no longer matched to the bowtie filter. Using fixed scanner settings, we measured how such vertical offsets change image noise and the dose at given points, and if the extra noise can be compensated by Adaptive Statistical Iterative Reconstruction (ASIR), which offsets the extra noise. On a GE Optima CT (120 kVp, 200 mA, no tube current modulation), a Catphan® 500 phantom was imaged at eleven heights between −5.75 and +14 cm and reconstructed with a standard and a soft kernel, using either filtered back projection (FBP) or ASIR. A pencil chamber gave point doses at five heights below the isocenter in a PMMA body phantom of 32 cm diameter. For the standard kernel with FBP, the SD was 11.01 HU at the centre and 14.01 HU at +14 cm (+27.2%). ASIR cut the noise by 35–40% at all heights, but at +14 cm it was still 33.9% higher than at the isocenter. At −8 cm the posterior dose was 47.9% higher and the anterior dose 27.9% lower, while the weighted CTDI changed by less than ±5%. ASIR therefore reduces, without removing, the effect of poor centering, and a weighted CTDI can hide large local dose differences.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Localiser-Based Software-Assisted Patient Positioning in Chest CT: A Propensity Score-Matched Study.Diagnostics (Basel, Switzerland) · 2026
- Optimization of image quality and radiation dose in low-dose pediatric chest CT Using 80 kVp acquisition and variable ASiR-V reconstruction: A retrospective comparative study.Medicine · 2026
- Dose optimization and modeled lifetime attributable cancer risk in brain CT using hybrid statistical iterative reconstructionEgyptian Journal of Radiology and Nuclear Medicine · 2026
- Evaluation of Image Noise from Computed Tomography in an Organ-Specific Anthropomorphic PhantomarXiv · 2026
- Attenuation Coefficient Imaging Using Regularization by Denoising.IEEE transactions on ultrasonics · 2026
- Attenuation Pre-Correction in Iterative PET Image Reconstruction: What is the Statistical Cost?IEEE transactions on radiation and plasma medical sciences · 2026