Patient Posture-Adaptive Non-Contact Respiratory Rate Measurement Inside a Moving Ambulance Using RGB and Thermal Cameras.
Daiga Nemoto, Nozomi Takahashi, Fumitaka Mukouyama, Taka-Aki Nakada, Yukihiro Nomura, Toshiya Nakaguchi
Department of Medical Engineering, Graduate School of Science and Engineering, Chiba University, 1-33, Yayoicho, Inage-ku, Chiba-shi 263-8522, Chiba, Japan. · Department of Emergency and Critical Care Medicine, Graduate School of Medicine, Chiba University, 1-8-1, Inohana, Chuo-ku, Chiba-shi 260-8670, Chiba, Japan.
Key points
From the paper's conclusion- We addressed this by patient posture-adaptive automatic image registration between RGB and thermal cameras and developing a nostril detection method robust to posture changes.
Abstract
Background/Objectives: Monitoring and evaluating respiratory rate (RR) plays a critical role during emergency transport. Conventional RR measurement methods have primarily relied on electrocardiogram-based approaches, which are susceptible to large errors caused by patient body movement and impose a burden on patients due to their contact-based nature. This study aimed to develop a non-contact RR measurement method using RGB and thermal cameras that can be applied inside a moving ambulance. Methods: One of the key challenges in non-contact RR measurement inside an ambulance is accommodating changes in patient posture. We addressed this by patient posture-adaptive automatic image registration between RGB and thermal cameras and developing a nostril detection method robust to posture changes. Results: The image registration error of the proposed method was evaluated at various distances from the camera, and the error was confirmed to remain within 1 pixel. RR measurement experiments simulating various respiratory conditions were then conducted inside a moving ambulance. The average of mean absolute error (MAE) was 0.95, 0.59, and 2.80 breaths per minute (breaths/min) for bradypnea, normal, and tachypnea, respectively, in the supine position, and 1.08, 1.90, and 4.38 breaths/min, respectively, in the semi-sitting position. Evaluation of the proposed nostril detection method further confirmed improved RR estimation accuracy compared with the conventional method. Conclusions: This study achieved patient posture-adaptive automatic image registration and robust nostril detection and successfully conducted experiments under realistic and challenging conditions, representing a substantial step toward the real-world implementation of a non-contact RR measurement system in ambulances.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Feasibility of Thermal Imaging for Contactless Respiratory Monitoring in Children.Acta paediatrica (Oslo, Norway : 1992) · 2026
- A Real-Time Low-Cost Platform for Non-Contact Respiratory Rate Monitoring: A Comparative Evaluation of RGB Optical Flow and Time-of-Flight Depth SensingarXiv · 2026
- Contactless Vital Sign Monitoring in Emergency Settings: A Factorial Study of Camera Position and Motion Using rPPGAlgorithms · 2026
- Colour-Channel Selection for Video-Based Respiratory Rate Estimation: A Diagnostic Study with a Single-Shot Held-Out Evaluation.Sensors (Basel, Switzerland) · 2026
- DeepRespNet: a hybrid attention-recurrent framework for non-contact respiratory rate estimationFrontiers in Physiology · 2026
- Camera-based respiratory rate monitoring in unmodified neonatal intensive care - a cross-sectional study in a UK hospital.EClinicalMedicine · 2026