Evaluation of the diagnostic validity of the HIV VISITECT CD4 point-of-care rapid diagnostic test in Uganda.
Oliver Nakamaanya, Bernard Bagaya, Willy Ssengooba, Erick Thokerunga, Henry Zamarano
Department of Immunology and Molecular Biology, Makerere University, Kampala, Uganda. olivecurtis1@gmail.com. · Department of Immunology and Molecular Biology, Makerere University, Kampala, Uganda.
Key points
From the paper's conclusion- The VISITECT CD4 LFA performed reliably as an AHD screening tool relative to the Pima Analyzer in resource-limited settings, particularly where instrument-based platforms are unavailable. It should complement, not replace, confirmatory testing; improvements in sensitivity and PPV remain warranted.
Abstract
BACKGROUND: HIV remains highly prevalent in Uganda, where CD4 T-cell counts are central to identifying and managing Advanced HIV Disease (AHD). Instrument-based point-of-care (POC) CD4 platforms face cost and operational constraints in resource-limited settings. The VISITECT CD4 lateral flow immunoassay (LFA) offers a non-instrument-based alternative, but its diagnostic validity against the Pima Analyzer had not been evaluated in Uganda. OBJECTIVES: To compare diagnostic validity of the VISITECT CD4 LFA against the Pima Analyzer at the 200 cells/µL AHD threshold and explore factors influencing performance. METHODS: A comparative cross-sectional study was conducted between June and August 2023 at three Kampala health centres. Venous blood from 351 HIV-positive adults was tested simultaneously on both platforms. Sensitivity, specificity, PPV, NPV, and likelihood ratios were calculated with 95% CIs using the Pima Analyzer-a point-of-care, not gold-standard, platform-as reference; estimates reflect relative, not absolute, performance. RESULTS: Most participants were female (68.11%), median age 31 years; mean CD4 count was 482 cells/µL (± 283.9) and AHD prevalence 22.79%. VISITECT CD4 LFA showed sensitivity of 86.25% (95% CI: 76.73-92.93), specificity of 90.41% (95% CI: 86.26-93.64), PPV of 72.63% (95% CI: 64.57-79.44), NPV of 95.70% (95% CI: 92.78-97.48), and inter-rater agreement κ = 0.957. No assessed factor significantly modified performance. CONCLUSIONS: The VISITECT CD4 LFA performed reliably as an AHD screening tool relative to the Pima Analyzer in resource-limited settings, particularly where instrument-based platforms are unavailable. It should complement, not replace, confirmatory testing; improvements in sensitivity and PPV remain warranted.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Performance Evaluation of the VISITECT CD4 Lateral Flow Point-Of-Care Assay for Identifying Advanced HIV Disease.Journal of clinical laboratory analysis · 2026
- Diagnostic accuracy of the VISITECT CD4 lateral-flow assay for screening advanced HIV disease: a systematic review and meta-analysis.International journal of infectious diseases : IJID : official publication of the International Society for Infectious Diseases · 2026
- Concordance Between Fingerprick and Venous Whole Blood CD4 Counts Using the Visitect CD4 Advanced disease platform.The Journal of infectious diseases · 2026
- Field performance evaluation of the GenBody ConfiSign HIV self-test for blood-based HIV screening in Uganda.PloS one · 2026
- Validation and performance of a rapid HIV-1 viral load assay.Microbiology spectrum · 2026
- Analytical Agreement of the GeneXpert HIV-1 Viral Load Assay Compared with the Abbott Alinity m HIV-1 Assay.Diagnostics (Basel, Switzerland) · 2026