Patient Radiation Dose Indices and Preliminary Local Diagnostic Reference Levels in Preoperative Intracranial Tumour Embolization Using a Low-Dose Protocol: A Single-Centre Study
Mariusz Stanisław Sowa, Joanna Sowa, Maciej Budzanowski
Key points
From the paper's conclusion- This outcome reflects a small, single-center cohort treated by a single operator using a low-dose protocol without 3D-RA.
Abstract
Background: Preoperative embolization of intracranial tumors, primarily performed for large, highly vascular meningiomas, is not recommended as a routine practice, as its clinical benefits have not been conclusively established. Procedure-specific diagnostic reference levels (DRLs) remain undefined in Poland and most European countries. This study aimed to establish local typical values and preliminary local DRLs for preoperative intracranial tumor embolization, with a separate analysis of a meningioma subgroup. Materials and Methods: A retrospective, single-center analysis was conducted on 52 consecutive embolization procedures performed between 2023 and 2026, of which 38 were meningiomas (73.1%). All procedures utilized a Philips Azurion ClarityIQ biplane system with a low-dose protocol (digital subtraction angiography [DSA] 2 frames per second [fps]; fluoroscopy 3.125 fps) with no three-dimensional rotational angiography (3D-RA) acquisitions. Dose–area product (DAP), reference air kerma (Ka,r), fluoroscopy time (FT), and the number of images were extracted from system-generated dose reports. Typical values were defined as the median (P50), and provisional local DRLs as the 75th percentile (P75). Vascular supply was classified according to the criteria of Richter and Schachenmayr. Results: For the entire cohort, the P50 and P75 values were 43.2 and 55.5 Gy·cm2 for DAP; 329.5 and 420.2 mGy for Ka,r; 1344.5 and 1809 s for FT; and 418 and 562.2 images, respectively. Exclusive or predominant external carotid artery supply was identified in 35 of 52 procedures type I and II (67.3%). Complete embolization was achieved in 27 of 52 procedures (51.9%). No complications were recorded. Conclusions: The values obtained were several times lower than Japanese national DRLs and historical Radiation Dose in Interventional Radiology (RAD-IR) values. This outcome reflects a small, single-center cohort treated by a single operator using a low-dose protocol without 3D-RA. These values may provide a local benchmark for dose optimization. However, establishing national DRLs for this procedure will require multicenter data.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Radiation Dose Metrics and Local Diagnostic Reference Levels in Low-Dose Stent-Assisted Coiling of Intracranial AneurysmsJournal of Clinical Medicine · 2026
- Real-time dosimetry in interventional radiology—comparing the operators’ radiation dose and associated factors for five types of transarterial embolization proceduresFrontiers in Public Health · 2026
- Local Diagnostic Reference Levels and Objective Image Quality in Mechanical Thrombectomy Performed with a Low-Dose Protocol: A Single-Center Observational Study.Journal of clinical medicine · 2026
- Patient radiation dose in diagnostic cerebral angiography for intracranial aneurysm: local diagnostic reference levels from a single-center study in Thailand.Radiation protection dosimetry · 2026
- Interventional radiology of the head and neck: Superselective intra-arterial chemoradiotherapy, preoperative tumor embolization, and emergency hemostasis.Auris, nasus, larynx · 2026
- Prospective Evaluation of Standalone Endovascular Embolization for Growing Meningiomas: A Multicenter Study Protocol.Neurologia medico-chirurgica · 2026