Decision Limits for Fetal Speckle-Tracking Outputs: Minimal Detectable Change and Threshold Instability in FetalHQ.
Jakub Mlodawski, Anna Zmelonek-Znamirowska, Lukasz Pawlik, Marta Mlodawska, Grzegorz Swiercz
Institute of Medical Sciences, Collegium Medicum, Jan Kochanowski University in Kielce, 25-516 Kielce, Poland. · Department of Information Systems, Kielce University of Technology, 25-314 Kielce, Poland.
Key points
From the paper's conclusion- The aim of this study was to evaluate the decision limits of selected FetalHQ parameters, including the minimal detectable change (MDC95), threshold instability, and the theoretical effect of averaging two independent analyses.
Abstract
Background: Fetal Heart Quantification (FetalHQ) enables detailed assessment of fetal cardiac morphology and function using two-dimensional speckle-tracking echocardiography. However, previous studies have focused primarily on conventional reproducibility metrics such as the intraclass correlation coefficient (ICC). From a clinical perspective, it is more relevant to determine what magnitude of change exceeds measurement noise and whether a single result may lead to unstable classification relative to normative thresholds. The aim of this study was to evaluate the decision limits of selected FetalHQ parameters, including the minimal detectable change (MDC95), threshold instability, and the theoretical effect of averaging two independent analyses. Materials and Methods: This methodological study included 93 paired analyses (_1 and _2) obtained from the same recorded four-chamber cine loop under the assumption of zero biological variability. The intra-operator subset (n = 34) and inter-operator subset (n = 59) were analyzed separately. For key parameters, we determined the proportion of usable pairs after quality filtering, bias, Bland-Altman 95% limits of agreement, MDC95, classification flip rates for normalized outputs, and the width of the uncertainty zone (grey zone) around diagnostic thresholds. In addition, we estimated the effect of a mean-of-two workflow on measurement error reduction. Results: The best analytical performance was observed for left ventricular volumetric parameters, particularly LV EDV and LV ESV, which showed the highest proportion of usable pairs (88.2-94.9%) and relatively low MDC95 values. Functional and deformation metrics performed substantially worse, including LV EF, LV FRAC, LV GLS, and RV TAPSE, all of which were characterized by lower usability and wider limits of agreement, especially in the inter-operator analysis. Classification instability was greatest for normalized parameters. In the inter-operator analysis, classification flips occurred in 33.3% of cases for LV EF centile <5, 31.4% for RV TAPSE lateral centile <5, 35.7% for LV GLS centile >95, and 39.2% for LV GLS Z-score >2. Under the theoretical assumption of independent random measurement error, the estimated MDC95 for the mean of two analyses was approximately 29% lower than that for a single analysis; this reduction was mathematically derived rather than prospectively observed. Conclusions: FetalHQ showed the highest analytical stability for simpler LV volumetric parameters, whereas functional and deformation outputs remained substantially more susceptible to measurement noise and classification instability, particularly between operators. In practice, this means that a single measurement obtained near normative thresholds may be insufficient for reliable interpretation. A theoretical mean-of-two approach may reduce random measurement uncertainty, but this strategy was not prospectively validated and does not overcome the limitations of the least analytically stable parameters.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Qualitätssicherung und Reproduzierbarkeit der fetalen 2D-Speckle-Tracking-EchokardiographieDie Gynäkologie · 2026
- Evaluation of Fetal Heart Quantification Analysis in Healthy Pregnancies Across Gestational Weeks. A Prospective Observational Study.Journal of gynecology obstetrics and human reproduction · 2026
- Reproducibility of fetal global longitudinal strain measured with speckle tracking echocardiography using a fixed cardiac cyclePLoS ONE · 2026
- Reference Ranges for Fetal Ventricular Global Longitudinal Strain (GLS) Using Bidimensional Speckle-Tracking Echocardiography: A Systematic Review.Journal of clinical medicine · 2026
- Addressing Key Limitations of Diastolic Function Assessment in Mouse Echocardiography by Enabling Robust Retrospective Analysis From a Standard Imaging ViewActa Physiologica · 2026
- Comparison of visual estimation of ventricular systolic function with quantitative echocardiographic markers in newborn infantsEcho Research and Practice · 2026