Initial Force Expression of Different Clear Aligner Systems in a Simulated Maxillary Canine Distalization Model with Lateral Incisor Agenesis: An In Vitro Comparative Study.

Alessio Verdecchia, Samuele Avolese, Andrea Tancredi Lugas, Cristina Bignardi, Mara Terzini, Claudia Dettori, Enrico Spinas, Teresa Cobo,

Orthodontics Division, Instituto Asturiano de Odontología, Universidad de Oviedo, 33006 Oviedo, Spain. · Specialization School in Orthodontics, Department of Surgical Sciences, Dental School of the University of Torino, 10129 Torino, Italy.

Preclinical

Key points

From the paper's conclusion
  • Background: This in vitro study compared initial force release by different clear aligner systems during simulated maxillary canine distalization with lateral incisor agenesis and evaluated attachment effects.

Abstract

BACKGROUND: This in vitro study compared initial force release by different clear aligner systems during simulated maxillary canine distalization with lateral incisor agenesis and evaluated attachment effects. METHODS: Five aligner systems were tested: SmartTrackTM (Invisalign®, Align Technology Inc., Santa Clara, CA, USA), TruGENTM (Spark®, Ormco Corporation, Brea, CA, USA), Essix® (SureSmile®, Dentsply Sirona Inc., Charlotte, NC, USA), MasterControl S (Angel Aligner®, Angelalign Technology Inc., Shanghai, China), and a directly 3D-printed aligner fabricated with TC-85 DAC resin (Graphy®, Graphy Inc., Seoul, Republic of Korea). Ten identical aligners per configuration were tested for 1 h at 37 °C in physiological solution. The ElectroForce® system measured the translational force at the canine's root-level support. Data were analyzed using descriptive statistics and two-way ANOVA. RESULTS: TruGENTM with attachments generated the highest mean force (1.25 N), followed by SmartTrackTM (0.61 N), Essix® (0.42 N), MasterControl S (0.39 N), and TC-85 DAC resin (0.01 N). Without attachments, forces were minimal for all systems. Force decay was observed in groups expressing measurable forces. CONCLUSIONS: Attachments enabled measurable translational force components in thermoformed systems. Force expression differed among systems, while TC-85 DAC resin produced negligible forces with or without attachments under experimental conditions. Further clinical validation is warranted.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Initial Force Expression of Different Clear Aligner Systems in a Simulated Maxillary Canine Distalization Model with Lateral Incisor Agenesis: An In Vitro Comparative Study — Science Skim