In Vivo Comparison of the Trueness of Conventional and Digital Interocclusal Registration.
Felix Förtsch, Johannes Ries, Tom Kasperlik, Manfred Wichmann, Ragai Edward Matta
Department of Prosthodontics, Erlangen University Hospital, Friedrich-Alexander-University Erlangen-Nuremberg, Glueckstrasse 11, 91054 Erlangen, Germany.
Key points
From the paper's conclusion- Virtual interocclusal registration showed greater trueness than the conventional workflow with a physical record, whereas manual articulation performed comparably to digital workflows.
Abstract
The objective was to evaluate the trueness of interocclusal registration obtained with three intraoral scanning systems versus a conventional impression and bite registration in fully dentate patients. Twenty adults with complete dentitions and Angle Class I occlusion were enrolled. Complete-arch scans were obtained with Primescan, iTero Element, and TRIOS 4, followed by bilateral virtual interocclusal registration. Conventional impressions and polyvinyl siloxane records were used to fabricate stone casts, digitized with a laboratory scanner with the record and after manual articulation in maximum intercuspation. Industrial scanner datasets served as participant-specific references. Three-dimensional deviations were analyzed by repeated-measures ANOVA and Holm-adjusted paired t-tests (α = 0.05). The conventional workflow with an interocclusal record showed the greatest deviation (797 ± 308 μm; p < 0.001). Manual articulation (331 ± 167 μm) did not differ significantly from Primescan (262 ± 120 μm), iTero Element (256 ± 101 μm), or TRIOS 4 (325 ± 168 μm) (p ≥ 0.223); no differences were found among scanners under the conditions of this study (p ≥ 0.468). Virtual interocclusal registration showed greater trueness than the conventional workflow with a physical record, whereas manual articulation performed comparably to digital workflows. These conclusions apply only to fully dentate patients with stable occlusal support and maximum intercuspation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Relative accuracy of intraoral photogrammetry and intraoral scanning with conventional scan-body for full-arch implants.Journal of prosthodontics : official journal of the American College of Prosthodontists · 2026
- Trueness and Precision of Three Novel Scan Body Designs with Horizontal Extensions for Complete-Arch Implant Impressions: A Comparative In Vitro Study with Intraoral Scanning and Photogrammetry Systems.Journal of functional biomaterials · 2026
- Record Extent and Occlusal Support in Scan-Assisted Interocclusal Registration: A Three-Dimensional In Vitro Study.Bioengineering (Basel, Switzerland) · 2026
- Accuracy of complete-arch implant impressions obtained with intraoral scanning, calibrated intraoral scan protocol, extraoral and intraoral photogrammetry, and conventional impression techniques: An in vitro study.Journal of prosthodontics : official journal of the American College of Prosthodontists · 2026
- Trueness and precision of 5 intraoral scanners under different conditions in arches: A 3D comparative in vitro analysis.Dental and medical problems · 2026
- Trueness of extraoral photogrammetry, intraoral photogrammetry and direct intraoral scanning in the full-arch digital implant impression: a comparative in vitro studyJournal of Dentistry · 2025