Laser-Fabricated APTES-Functionalized Interdigitated Electrodes with a Portable Readout: A Proof-of-Concept Impedimetric Response to Aqueous Nosema Spore Suspensions.
Kamil Çınar, Christopher Mayack, Hülya Eraslan-Gültekin, Ali Fuat Ergenç, Melike Kefeli, Asiye Gök Yurttaş
Faculty of Science, Department of Physics, Gebze Technical University, 41400 Kocaeli, Türkiye. · Department of Biology, College of Science and Health, William Paterson University, Wayne, NJ 07470, USA.
Key points
From the paper's conclusion- A self-contained readout built around an ESP32 microcontroller reproduced the direction of this change, its output voltage rising by approximately 33% after the suspension was applied.
Abstract
Nosemosis, caused by the microsporidian parasites Nosema apis and Nosema ceranae, is a major contributor to honeybee (Apis mellifera) colony losses, yet its diagnosis still relies on destructive sampling followed by light microscopy or nucleic-acid amplification in a laboratory. We describe an impedimetric transducer intended as a first step toward a non-destructive alternative. Interdigitated gold electrodes were patterned on standard microscope slides by nanosecond laser ablation, a route that requires neither photolithography nor cleanroom processing, and were functionalized with (3-aminopropyl)triethoxysilane (APTES) deposited from either ethanol or toluene. Because the 0.01-50 MHz measurement window lies far above the range in which faradaic charge transfer contributes measurably, the response is non-faradaic and is governed by the permittivity and the ionic conductivity of the deposited droplet rather than by interfacial charge transfer. Films deposited from ethanol gave the largest and most reproducible contrast between an aqueous spore suspension (1500 spores/µL) and a deionized-water control. At the structural resonance of the electrode (≈16 MHz) the contrast in impedance magnitude is approximately 10% and is not resolved above the chip-to-chip scatter, which is expected because the magnitude at resonance is set by the series resistance of the electrode path rather than by the droplet. The two liquids are instead separated by the reactive component: the imaginary part of the impedance differs by 15.7 ± 1.8 MΩ near 6.8 MHz, about nine times the replicate uncertainty, and the phase by 33 ± 7° at 4.5 MHz. A self-contained readout built around an ESP32 microcontroller reproduced the direction of this change, its output voltage rising by approximately 33% after the suspension was applied. The scope of these observations is deliberately narrow: a single spore concentration in deionized water, no spore-free filtrate control, no calibration series, no selectivity panel and no measurement in a fecal matrix. The contribution is accordingly methodological-a cleanroom-free fabrication route, a quantified solvent dependence of the APTES film, and the transfer of the measurement from a benchtop spectrometer to a portable unit-rather than a validated diagnostic assay.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- 3D focused impedance spectroscopy for label-free profiling of thermal stress in microalgae.Lab on a chip · 2026
- COMSOL-Based Impedance and Sensitivity Analysis of an Interdigitated Electrode Biosensor for Dengue NS1 Detection.Biosensors · 2026
- Dielectrophoretic Enrichment Coupled with Impedance Spectroscopy for Real-Time Bacterial Detection and Antibiotic Susceptibility Testing Using an Interdigitated Wave Electrode Array.Sensors (Basel, Switzerland) · 2026
- Sustainable Microelectrode Fabrication for Electrochemical Impedance-Based In Situ Monitoring In Vitro Models.ACS applied bio materials · 2026
- Mesoporous PU/PEDOT:PSS Electrodes Reveal Population-Level Stochastic Bioelectrical Dynamics in Marine Diatoms.Advanced science (Weinheim, Baden-Wurttemberg, Germany) · 2026
- Electrochemical Aptasensing Platform for Culture-Free, Selective Detection of Airborne Bacterial SpeciesACS Sensors · 2026