Optimisation of Q.Clear (BSREM) Reconstruction Parameters for 123I and 177Lu Imaging with a 360° CZT Detector.
Rafaela Carvalhais, Joana Teixeira, Vera Antunes, João Santos
Faculty of Sciences, University of Porto, 4169-007 Porto, Portugal. · Medical Physics, Radiobiology and Radiation Protection Group, IPO Porto Research Center (CI-IPOP), Portuguese Institute of Oncology of Porto (IPO Porto), 4200-072 Porto, Portugal.
Key points
From the paper's conclusion- This study aimed to optimise Q.Clear reconstruction parameters for 123I and 177Lu imaging.
Abstract
Accurate quantitative SPECT depends strongly on the reconstruction protocol, particularly for penalised-likelihood reconstruction algorithms, such as Q.Clear (BSREM). This study aimed to optimise Q.Clear reconstruction parameters for 123I and 177Lu imaging. SPECT images of a NEMA IEC PET Body phantom filled with 123I and 177Lu were acquired on a ring-shaped CZT camera and on a conventional Anger camera. The images were reconstructed using Q.Clear and OSEM and evaluated using quantitative image-quality metrics. Q.Clear was optimised for each radionuclide. The 123I dataset was reconstructed using different combinations of the regularisation parameters β and γ. The 177Lu dataset was reconstructed using literature-reported parameters, and the best-performing reconstruction was fine-tuned. The best-performing Q.Clear reconstructions were compared with corresponding OSEM reconstructions. Overall, increasing β reduced the recovery coefficients, background variability, and contrast recovery coefficients while increasing the signal-to-noise ratio. γ had no appreciable influence on the metrics within the investigated range. The candidate optimised StarGuide/Q.Clear reconstructions outperformed the standard Symbia/OSEM reconstructions in terms of activity recovery, whereas Symbia/OSEM produced lower image noise for most of the analysed VOIs. This study identified settings associated with improved quantitative metrics within the evaluated parameter space, with potential implications for quantitative imaging requiring clinical validation. Nevertheless, multiple reconstruction parameter combinations remain unexplored.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Systematic evaluation of energy window and reconstruction parameters for quantitative lutetium‐177 SPECT imaging on a CZT‐based starguide SPECT/CT systemJournal of Applied Clinical Medical Physics · 2026
- Optimisation of reconstruction parameters for lutetium-177 SPECT imaging: comparative performance of cadmium-zinc-telluride and NaI(Tl) detectors.Nuclear medicine communications · 2026
- Quantitative Reconstructions for Renal Sestamibi SPECT/CT: A Modeled Comparison of 360-Degree Versus 180-Degree Acquisitions.Nuklearmedizin. Nuclear medicine · 2026
- Impact of CT parameter reduction on quantitative SPECT/CT assessment in 99mTc imaging: A phantom study.Radiography (London, England : 1995) · 2026
- Experimental hybrid spectral CT with Cramér-Rao lower bound-optimized weighting for quantitative iodine imaging.medRxiv : the preprint server for health sciences · 2026
- Application of Deep Learning Algorithm-Based Image Reconstruction in Improving Abdominal CT Image Quality in Adrenal and Renal Lesions Detection.Radiology research and practice · 2026