Optimisation of reconstruction parameters for lutetium-177 SPECT imaging: comparative performance of cadmium-zinc-telluride and NaI(Tl) detectors.
Hongjin Zhou, Shen Wang, Wei Zheng, Ruyi Zhang, Qiang Jia
Department of Nuclear Medicine, Tianjin Medical University General Hospital, Tianjin, China.
Key points
From the paper's conclusion- For the 177Lu 113 keV peak, CZT detectors markedly enhance SPECT imaging, improving lesion detectability and quantification. Tailored reconstruction optimization further strengthens image quality and reliability, supporting precision dosimetry and CZT adoption in radionuclide therapy.
Abstract
BACKGROUND: Lutetium-177 (177Lu) is a promising theranostic radionuclide. Accurate quantitative SPECT imaging is essential for 177Lu patient-specific dosimetry. Although the 113 keV photopeak has a higher scatter fraction, it matches the high energy resolution of cadmium-zinc-telluride (CZT) detectors and avoids 208 keV limitations. However, whether 113 keV-only imaging provides sufficient lesion detectability remains unclear. Focusing on the 113 keV photopeak, this study compared two dual-head SPECT/computed tomography systems: conventional NaI(Tl) scintillators with low-energy high-resolution collimators, and CZT semiconductor detectors with wide-energy high-resolution collimators. This study compares these technologies and establishes optimized reconstruction protocols. METHODS: National Electrical Manufacturers Association/International Electrotechnical Commission phantoms were scanned on both systems using the 113 keV photopeak. Ordered subset expectation maximization reconstructions were performed (2-12 iterations) with Gaussian [full-width-at-half-maximum (FWHM): 1.5-4.0 mm) or Butterworth filters (cut-off frequency: 0.25-0.50 cycle/pixel). Attenuation, scatter, and resolution corrections were applied. Image quality was quantified using background coefficient of variation (COV_B), contrast recovery coefficient (CRC), contrast-to-noise ratio (CNR), and signal-to-noise ratio (SNR). RESULTS: CZT delivered superior image quality, achieving higher CRC, CNR, and SNR with lower COV_B than NaI(Tl). Optimal parameters were 2 iterations, 10 subsets, Gaussian FWHM 3.0 mm, and Butterworth cut-off 0.35 cycles/pixel (CZT) and 0.30 cycles/pixel [NaI(Tl)]. These settings yielded robust performance across target-to-background ratios of 8 : 1, 16 : 1, and 32 : 1. CONCLUSION: For the 177Lu 113 keV peak, CZT detectors markedly enhance SPECT imaging, improving lesion detectability and quantification. Tailored reconstruction optimization further strengthens image quality and reliability, supporting precision dosimetry and CZT adoption in radionuclide therapy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Systematic evaluation of energy window and reconstruction parameters for quantitative lutetium‐177 SPECT imaging on a CZT‐based starguide SPECT/CT systemJournal of Applied Clinical Medical Physics · 2026
- Optimisation of Q.Clear (BSREM) Reconstruction Parameters for 123I and 177Lu Imaging with a 360° CZT Detector.Bioengineering (Basel, Switzerland) · 2026
- Clinical evaluation of Chang attenuation-correction parameterization for CT-free 177Lu SPECT organ quantification: implications for safety dosimetry.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Experimental Evaluation of DE-SPECT: A Hyperspectral SPECT System for Region-Selective 3-D Gamma-Ray Spectroscopy.IEEE transactions on radiation and plasma medical sciences · 2026
- Quantitative Comparison of SPECT and PET Performance for Clinical Theranostic ApplicationsJournal of Nuclear Medicine · 2025
- Comprehensive acceptance testing and quality control of Single Photon Emission Computed Tomography/Computed Tomography systems.Nuclear medicine communications · 2026