Quantitative Reconstructions for Renal Sestamibi SPECT/CT: A Modeled Comparison of 360-Degree Versus 180-Degree Acquisitions.
Jorge D Oldan, Jennifer A Schroeder, Mathurika Jeyasingam, Lilja B Solnes, Emi Lou Mondragon, Thad Benefield, Mohammad E Allaf, Mark Tulchinsky,
University of North Carolina, Division of Molecular Imaging and Therapeutics, Department of Radiology, NC, USA, Chapel Hill. · SUNY Upstate University Hospital, Department of Interventional Radiology, NY, USA, Syracuse.
Key points
From the paper's conclusion- Modeled 180-degree reconstructions of quantitative SPECT/CT are qualitatively inferior to quantitative reconstructions of 360-degree acquisitions, although there was no difference in the diagnostic accuracy between the reconstruction methods. These results suggest that 180-degree acquisitions are adequate for diagnosis.
Abstract
RATIONALE: 99m Tc-sestamibi single-photon emission computed tomography/computed tomography (SPECT/CT) is a cost-effective and accurate means of characterizing indeterminate renal masses. This is scant literature on 99m Tc-sestamibi protocol optimization. METHODS: We investigated the difference between modeled, quantitative 180-degree and 360-degree acquisitions for 99mTc-sestamibi SPECT/CT in a cohort of 50 patients who were prospectively enrolled on a scan-and-resect protocol. Two experienced, dual-boarded observers rated each renal mass for image quality (1-5 scale) and diagnosis (benign to malignant, 1-5 scale) in a blinded, randomized order. Ordinal logistic regression was used, incorporating random intercepts for patient and reader; inter-observer variability was assessed. RESULTS: Image quality ratings were significantly higher for 360-degree acquisitions versus 180-degree (P<0.0001; odds ratio 0.22, 95% CI 0.12-0.39). The odds ratio was 0.22 (95% CI: 0.12, 0.39). There was insufficient evidence that the modeled angle of acquisition was associated with scan positivity (P=0.17). The odds ratio was 0.67 (95% CI: 0.37, 1.19). Diagnostic positivity ratings did not differ between reconstruction angles (P=0.17; odds ratio 0.67, 95% CI 0.37-1.19). CONCLUSIONS: Modeled 180-degree reconstructions of quantitative SPECT/CT are qualitatively inferior to quantitative reconstructions of 360-degree acquisitions, although there was no difference in the diagnostic accuracy between the reconstruction methods. These results suggest that 180-degree acquisitions are adequate for diagnosis.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Optimisation of Q.Clear (BSREM) Reconstruction Parameters for 123I and 177Lu Imaging with a 360° CZT Detector.Bioengineering (Basel, Switzerland) · 2026
- Practical Recoverability and Numerical Identifiability of Kinetic Rate Constants from Noisy Time-Activity Curves in a Patient-Derived Renogram Model.Molecular imaging · 2026
- Impact of ROI delineation methods on the repeatability of quantitative indices in 99mTc-MAG3 diuretic renography.The quarterly journal of nuclear medicine and molecular imaging : official publication of the Italian Association of Nuclear Medicine (AIMN) [and] the International Association of Radiopharmacology (IAR), [and] Section of the Society of... · 2026
- Optimization of spectral photon-counting CT reconstruction parameters for improved detection of peritoneal metastases compared with energy-integrating detector dual-energy CT: a phantom study.European radiology experimental · 2026
- Dose-aware 5-star image quality rating in CT: Findings from the IAEA-MGH study.European journal of radiology · 2026
- Quantitative Comparison of SPECT and PET Performance for Clinical Theranostic ApplicationsJournal of Nuclear Medicine · 2025