Traceable Calibration of Process Oscillation Type Density Meters: Corrosion Induced Drift Correction
Jintao Wang, Jingyue Zhang, Xiang Liu, Xu Chang, Xiaowen Su
Key points
From the paper's conclusion- Periodic, traceable full coefficient calibration is essential to correct corrosion induced drift, and preliminary full coefficient calibration intervals of 6 to 36 months follow from those instruments according to service severity.
Abstract
Oscillation type (vibrating tube) density meters are widely used for quality control and custody transfer in petrochemical processes. In service, corrosion and wear of the tube wall change the effective mass of the resonator. The measured density drifts, and the drift accumulates with time. A calibration methodology for process oscillation type density meters is used, traceable to the national primary density standard of the National Institute of Metrology (NIM) of China. The analytical relationship between the oscillation period and the liquid density was derived and extended with temperature and pressure correction terms, and a complete calibration system was established to reproduce pressure and temperature conditions. For three meters of two types, in service from lightly to severely corrosive duty, out of date coefficients produced a density error of 3.35 kg/m3 at the density of water, where the internal corrosion of the tube has its greatest effect. After a full coefficient calibration the residual at the reference liquids fell to 0.01–0.10 kg/m3, with an expanded uncertainty of 0.1 kg/m3 (k = 2). The correction model and the traceable procedure therefore restore the required accuracy. Periodic, traceable full coefficient calibration is essential to correct corrosion induced drift, and preliminary full coefficient calibration intervals of 6 to 36 months follow from those instruments according to service severity.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Determination of Reference Density from Measured Density and Temperature Data: A Diesel Fuel Case StudyProcesses · 2026
- Wide-Ranging Measurements of the Viscosity and Density of Neon with a Combined Visco-Densimeter and Reevaluated Results of an Oscillating-Disk ViscometerInternational Journal of Thermophysics · 2026
- A Unified Metrological Platform for Calibrating Emission Monitors: Quantifying and Correcting the Lab-to-Field Systematic BiasMeasurement Science and Technology · 2026
- Transfer of off-line calibration models to an in-line near-infrared high-speed monitoring system for tablet uniformity.Journal of pharmaceutical sciences · 2026
- From Potentiometric Titration to Electrometric Conductivity: A Critical Review and Metrological Comparison of ASTM D6470 and ASTM D3230 toward Method Replacement in Salt in Crude Analysis.Critical reviews in analytical chemistry · 2026
- Field-relevant metrological performance of water meters: a topical review for water-infrastructure monitoringMeasurement Science and Technology · 2026