Transfer of off-line calibration models to an in-line near-infrared high-speed monitoring system for tablet uniformity.
Cristian Kulcițki, Anna Novikova, Claudia S Leopold
Department of Pharmaceutical Technology, Institute of Pharmacy, University of Hamburg, Bundesstraße 45, D-20146 Hamburg, Germany. · Fette Compacting, Grabauer Straße 24, D-21493 Schwarzenbek, Germany.
Key points
From the paper's conclusion- Both, calibration transfer methods as well as the off-line calibration with tablet densities and temperatures similar to those during the in-line spectra acquisition reduced the systematic prediction bias successfully.
Abstract
In-line near-infrared (NIR) spectroscopy is a widely used method to monitor tableting processes. The acquisition of in-line NIR spectra of individual tablets is possible between their compression and ejection stages in the tablet press. It was shown, that in-line spectra used for the API content prediction with off-line calibration models led to a systematic prediction bias. To further investigate factors that may influence this bias, in the present study off-line calibration models were developed with tablets of different densities obtained by varying the compression force which were exposed to either room temperature or a range between 22.3 and 51.5°C during spectral acquisition. Univariate (linear response correction) and multivariate (direct orthogonalization) calibration transfer methods were investigated using off-line and in-line acquired calibration and test spectra with different API contents. The different densities of the calibration tablets and the temperature during calibration had a negligible influence on the calibration models and on the cross-validation results. However, both influencing factors had a major effect on the content prediction bias obtained with in-line spectra. Both, calibration transfer methods as well as the off-line calibration with tablet densities and temperatures similar to those during the in-line spectra acquisition reduced the systematic prediction bias successfully.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of the spectral information acquired by an in-line high-speed near-infrared reflection sensor in a tableting environment.European journal of pharmaceutical sciences : official journal of the European Federation for Pharmaceutical Sciences · 2026
- In-line laser triangulation-based 3D profiling for real-time tablet monitoring in a rotary tablet press.Journal of pharmaceutical sciences · 2026
- Evaluating near-infrared spectroscopy and multivariate modelling for predicting dissolution profiles of 3D printed extended-release tablets.International journal of pharmaceutics · 2026
- Conformal Coverage Collapse Under NIR Instrument Transfer and the Price of Restoring It with Labelled Target SamplesApplied Sciences · 2026
- Comparison and Internal Cross-Validation of Segmented and Conventional Calibration Models for a Near-Infrared Nephelometric Turbidity Sensor.Sensors (Basel, Switzerland) · 2026
- From Benchtop to Handheld Systems: Calibration Transfer for On-Site NIR System Butter Quality Monitoring.Sensors (Basel, Switzerland) · 2026