Wide-Ranging Measurements of the Viscosity and Density of Neon with a Combined Visco-Densimeter and Reevaluated Results of an Oscillating-Disk Viscometer
Ulrike Kochan-Eilers, Ruben Harten, Eckhard Vogel, Karsten Meier
Key points
From the paper's conclusion- Furthermore, measurements of the viscosity of dilute neon gas carried out by one of the authors with an oscillating-disk viscometer in 1984 were reevaluated by applying a highly accurate theoretical value for the viscosity of helium in the calibration of the viscometer.
Abstract
Accurate wide-ranging measurements of the viscosity and density of supercritical neon were performed with a combined vibrating-wire viscometer and single-sinker densimeter. They cover the temperature range between 300 K and 450 K with pressures up to 30 MPa. The viscosity measurements are characterized by a relative expanded (coverage factor $$k=2$$ k = 2 ) uncertainty of 0.25 %, while the relative expanded ( $$k=2$$ k = 2 ) uncertainty in the density measurements is 0.05 %, except in the low-density regime, where it increases up to 0.1 % at 5 kg/m $$^3$$ 3 . The vibrating-wire viscometer was calibrated with helium using a highly accurate theoretical value for the viscosity of helium in the limit of zero density at 293 K. To validate the calibration of the viscometer, measurements on nitrogen along the two isotherms 293 K and 423 K were carried out and compared with literature data, which confirmed the calibration. Our measurement results for the viscosity of neon agree with the most accurate literature data within their mutual uncertainties in the regions where they overlap but deviate by up to 1.1 % from the corresponding-states model of Huber [M. L. Huber, NIST Internal Report 8209, National Institute of Standards and Technology, 2018]. Our results for the density agree with the current reference equation of state for neon within the uncertainty of our data. Furthermore, measurements of the viscosity of dilute neon gas carried out by one of the authors with an oscillating-disk viscometer in 1984 were reevaluated by applying a highly accurate theoretical value for the viscosity of helium in the calibration of the viscometer. After the reevaluation, the relative expanded $$(k=2)$$ ( k = 2 ) uncertainty in these measurements amounts to 0.2 % to 0.4 %. Moreover, we present new correlations for the viscosity and thermal conductivity of neon in the limit of zero density, which are based on recent accurate ab initio calculations.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Reference Correlation for the Viscosity of Helium.International journal of thermophysics · 2026
- Viscosity Measurement of Liquid Propane, n -Butane and i -Butane at Temperatures from 135 to 225 K and Pressures up to 10 MPaJournal of Chemical & Engineering Data · 2026
- Measurements of Density and Viscosity for Methyl Dodecanoate and Ethyl Tetradecanoate and Speed of Sound in Ethyl Tetradecanoate at Temperatures from (298 to 423) K and Pressures from (0.1 to 100) MPaJournal of Chemical & Engineering Data · 2026
- Traceable Calibration of Process Oscillation Type Density Meters: Corrosion Induced Drift CorrectionApplied Sciences · 2026
- Coupling residual entropy scaling and revised Enskog theory (REST): A predictive framework for multicomponent transport properties from dilute gas to dense liquidThe Journal of Chemical Physics · 2026
- Measurements of Density and Viscosity for Binary (CO2 + O2) and Quaternary (CO2 + O2 + N2 + Ar) MixturesJournal of Chemical & Engineering Data · 2026