Key points
From the paper's conclusion- We evaluated the 15-year institutional transition from three-dimensional conformal radiotherapy (3D-CRT) to volumetric modulated arc therapy (VMAT) and compared their dosimetric performance in pediatric focal brain radiotherapy (FBRT).
Abstract
Background/objectives Minimizing radiation-induced late effects is critical in pediatric neuro-oncology, where long life expectancy is associated with treatment-related toxicity. We evaluated the 15-year institutional transition from three-dimensional conformal radiotherapy (3D-CRT) to volumetric modulated arc therapy (VMAT) and compared their dosimetric performance in pediatric focal brain radiotherapy (FBRT). Methods Between 2011 and 2025, 27,393 patients were treated, including 175 pediatric cases (≤18 years). FBRT cases were selected for dosimetric analysis because they are the most common pediatric brain irradiation at our institution and involve small-to-medium targets located near radiosensitive critical structures, enabling a sensitive comparison of technique-dependent dosimetric differences. FBRT plans from 27 patients were analyzed (3D-CRT: n = 13; VMAT: n = 14) using a unified script-based workflow in RayStation combined with a standardized AI-based auto-segmentation tool. Metrics included V95%, homogeneity index (HI), conformity index (CI), and high-dose spillage (V102%). Exploratory organ at risk (OAR) analysis was performed in a more restricted prescription uniform subgroup. In addition, overall survival (OS) was evaluated using available clinical follow-up data from FBRT patients. Results VMAT significantly reduced high-dose volumes (V102%: 0.01% vs. 0.78%, p = 3.8 × 10 −4 ), improved conformity (CI: 0.85 vs. 0.64, p = 0.008) and homogeneity (HI: 1.07 vs. 1.10, p = 0.048) compared with 3D-CRT. Target coverage was comparable. OAR analysis showed consistently reduced maximum doses to the optic chiasm and pituitary gland. No statistically significant difference in overall survival was observed between treatment techniques (p = 0.26). Conclusion Overall, our study demonstrates that the application of VMAT in pediatric focal brain irradiation improves the precision of dose distribution and may contribute to reducing long-term neurocognitive and endocrine toxicities. These findings support the clinical use of VMAT, with the potential to reduce late treatment-related toxicity in the pediatric patient population.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Unified Planning Platform Comparison of VMAT, Helical Tomotherapy, and IMRT in Nasopharyngeal Carcinoma: Dosimetric Superiority and Organ Sparing AnalysisTechnology in Cancer Research & Treatment · 2026
- RADT-49 Dosimetric Comparison of adjuvant Stereotactic Radiotherapy (SRT) Versus Conventionally Fractionated Volumetric Modulated Arc Therapy (VMAT) in Glioblastoma MultiformeNeuro-Oncology · 2026
- Evaluation of 3D-CRT vs. VMAT for radiotherapy of whole breast with simultaneous integrated boostActa Oncologica · 2026
- A Dosimetric and Biological Comparison of Volumetric Modulated Arc Therapy and Adaptive Aperture Proton Therapy for Hippocampal-Sparing Whole-Brain Radiotherapy.Technology in cancer research & treatment · 2026
- Dosimetric Comparison of Five Radiotherapy Techniques in Left-Sided Breast Cancer Patients Following Breast-Conserving Surgery.The breast journal · 2026
- RapidArc dynamic versus IMRT and VMAT for endometrial cancer SIB radiotherapy.Radiation oncology (London, England) · 2026