Confidence-gated temporal voting for EMG-based prosthetic arm control: a low-cost end-to-end pipeline with quantified rejection behaviour.

Anagha Shinde, Virendra Shete, Ninad Mehendale

School of Engineering and Science, MIT ADT Campus, MIT Art, Design and Technology University, Pune, Maharashtra, India. · K. J. Somaiya School of Engineering, Somaiya Vidyavihar University, Mumbai, India.

Computational

Key points

From the paper's conclusion
  • Conclusion The results support a benchtop engineering proof of concept with quantified rejection and latency behaviour, not clinical safety or deployment readiness.

Abstract

BACKGROUND: Surface electromyography (sEMG) is the dominant non-invasive route to volitional prosthetic control, but a closed-world classifier issues a command for every input, including inputs corresponding to no trained gesture. This study evaluated a low-cost, three-gesture sEMG control pipeline combining a Random Forest classifier with rejection on the maximum Random-Forest class probability and temporal majority voting, and measured the resulting rejection behaviour rather than assuming it. METHODS: Three data sources were analysed: modelled synthetic sEMG; Ninapro DB5 (10 subjects, 16 channels at 200 Hz, documented three-channel subset); and live three-channel sEMG from 13 healthy limb-intact adults recorded over three sessions with MyoWare sensors spaced equally around the forearm flexor group. Each of the 117 recordings lasted 120 s, comprising a 60 s continuous phase and four 10 s sustained contractions separated by 5 s of rest, yielding 21,294 windows of which 4,212 were rest. Channel interleaving gave an effective 333.3 Hz per channel. All datasets were segmented in physical time (0.6 s window, 0.2 s overlap) without resampling, and four time-domain features (MAV, RMS, WL, ZC) were classified by a 100-tree Random Forest. Evaluation used repetition-level, cross-session and leave-one-participant-out (LOPO) splits, with calibration and threshold selection confined to a participant-disjoint partition. Models were trained and tested within each domain, so the three accuracies index benchmark difficulty rather than transfer. RESULTS: LOPO accuracy was 80.2% (synthetic), 74.5% (DB5) and 80.2% (live). The pooled window-level split used originally overestimated live accuracy by 7.7 percentage points. Expected calibration error was 0.147 uncalibrated and 0.038 after Platt scaling. Open-set AUROC was 0.842 against rest, 0.784 under electrode displacement and 0.741 for held-out movements, and 7.4% of windows were confidently incorrect. The complete pipeline reduced false actuation from 19.80% to 5.00% and command switching from 0.16 to 0.02  s - 1 , while rejecting 18.40% of windows and raising median latency from 0.60 to 1.95 s. The prototype completed 12 of 15 benchtop trials. CONCLUSION: The results support a benchtop engineering proof of concept with quantified rejection and latency behaviour, not clinical safety or deployment readiness.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Confidence-gated temporal voting for EMG-based prosthetic arm control: a low-cost end-to-end pipeline with quantified rejection behaviour — Science Skim