Differences in novel Schistosoma mansoni Sm25 and Sm29 rapid diagnostic test performance based on sample matrix.
Madelaine Usey Talbot, KathyJean Farnam, Jean Saunders, E Brook Goodhew, Gabrielle Smith, Sylvia Ossai, Yong Wang, Sukwan Handali,
Laboratory Leadership Service, Division of Workforce Development, Public Health Infrastructure Center, Centers for Disease Control and Prevention, Atlanta, GA, United States. · Drugs and Diagnostics for Tropical Diseases, San Diego, CA, United States.
Key points
From the paper's conclusion- While both Sm25 and Sm29 RDTs exhibited high specificity and stability that surpassed TPP criteria, their sensitivity fell short of TPP goals. Sensitivity was highest with serum, followed by DBS, then whole blood, which would be the most useful for point-of-care settings. In addition, we identified a gap in the ability of currently used recombinant antigens to detect all people infected with S. mansoni, regardless of test platform.
Abstract
INTRODUCTION: Schistosomiasis is a neglected tropical disease caused by parasitic worms of the genus Schistosoma. As schistosomiasis control programs approach elimination as a public health problem, there is an urgent need for sensitive diagnostic assays that can easily be deployed in the field at low cost. For schistosomiasis transmission interruption and surveillance use cases, the diagnostic Target Product Profile (TPP) proposed a two-step testing strategy involving an initial point-of-care screening test capable of detecting either current or former infections with sensitivity > 0.95, specificity > 0.80, and result stability ≥ 0.5 hours. To meet this need, Drugs and Diagnostics for Tropical Diseases designed two novel rapid diagnostic test (RDT) prototypes that detect antibodies against recombinant Schistosoma mansoni antigens: Sm25 and Sm29. METHODS: We evaluated the performance of these RDT prototypes using serum, whole blood, and dried blood spots (DBS). RDT results were determined by two visual readers and a Lumos automated reader, which also provided a quantitative readout of test line intensity. Lumos results were used to quantify test line stability over time. Further, we compared RDT results to other serologic assays (ELISA and the multiplex bead assay) that detect antibodies against Sm25, Sm29 and two additional S. mansoni antigens. RESULTS: While both Sm25 and Sm29 RDTs exhibited high specificity and stability that surpassed TPP criteria, their sensitivity fell short of TPP goals. Sensitivity was highest with serum, followed by DBS, then whole blood, which would be the most useful for point-of-care settings. In addition, we identified a gap in the ability of currently used recombinant antigens to detect all people infected with S. mansoni, regardless of test platform. DISCUSSION: Our findings underscore that while these prototypes represent a promising start, there is still a case for novel antigen discovery and generation of RDTs with increased sensitivity to meet TPP criteria.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of six different tests for Schistosoma haematobium diagnosis in a near-elimination setting: A prospective observational diagnostic accuracy study.PLoS neglected tropical diseases · 2026
- Diagnostic accuracy of rapid and point-of-care tests for Schistosoma haematobium in African populations: a systematic review and meta-analysis.Frontiers in parasitology · 2026
- Laboratory evaluation of five rapid diagnostic tests for the serological diagnosis of chronic Trypanosoma cruzi infection in Yucatan, MexicomedRxiv · 2026
- Development and assessment of a standardized microscopy protocol for the diagnosis of male genital schistosomiasis by detection of Schistosoma haematobium eggs in human semenFrontiers in Parasitology · 2026
- Diagnostic Accuracy of the TR Chagas Bio-Manguinhos Rapid Test in the Municipality of Canguçu, Rio Grande do Sul, a Historically Endemic Area for Chagas Disease in Southern Brazil: a prospective study.Acta tropica · 2026
- Comparison of Cobas Plasma Separation Cards and Whatman filter paper cards versus conventional serum samples for anti-Trypansoma cruzi antibody detection.PLoS neglected tropical diseases · 2026