Laboratory evaluation of five rapid diagnostic tests for the serological diagnosis of chronic Trypanosoma cruzi infection in Yucatan, Mexico
R. Heslop, G. Pellegrin, P. Martinez-Vega, C. Teh-Poot, J. Torres-Montero, H. Noh-Pech, K. Dzul-Rosado, F. Puerto-Manzano,
Key points
From the paper's conclusion- Here, we evaluated the diagnostic performance of five RDTs: Chagas Rapid Test (Certum), Bio-Chagas Rapid Test (Mexlab), BiolineTM Chagas Ab (Abbott), Chagas Rapid Test (Amunet) and Chagas Sero K-SeT (Coris BioConcept).
Abstract
Chagas disease (CD), caused by Trypanosoma cruzi, is a significant public health challenge in Latin America, where an estimated 70% of infected individuals remain undiagnosed. Rapid diagnostic tests (RDTs) offer a promising solution for new screening strategies, yet their performance has not been systematically assessed in Mexican cohorts. Here, we evaluated the diagnostic performance of five RDTs: Chagas Rapid Test (Certum), Bio-Chagas Rapid Test (Mexlab), BiolineTM Chagas Ab (Abbott), Chagas Rapid Test (Amunet) and Chagas Sero K-SeT (Coris BioConcept). Performance was tested using 79 archived human serum samples from Yucatan. All tests were benchmarked against the diagnostic reference algorithm established by the Mexican Official Norm (NOM), where samples were characterised as positive if they were reactive to [≥] 2 conventional serological assays (Chagatest Lisado, Chagatest Recombinante v4.0 and Chagatest HAI). The RDTs demonstrated variable diagnostic sensitivity. The BiolineTM Chagas Ab test demonstrated the highest sensitivity (97.2%, 95% CI: 85.5 - 99.9; 35/36 positive samples detected) and almost perfect statistical agreement with the NOM reference algorithm ({kappa}= 0.90; p < 0.0001). Similarly, the Chagas Sero K-SeT prototype showed a high sensitivity (91.7%, 95% CI: 77.5 - 98.2; 33/36 positive samples detected) and substantial concordance ({kappa}= 0.87, p < 0.0001) with the NOM-based composite result. In contrast, the remaining RDTs demonstrated low sensitivities (Certum: 0.0%, 95% CI: 0.0-9.7; Mexlab: 5.6%, 95% CI: 0.7-18.7; Amunet: 33.3%, 95% CI: 18.6-51.0), limiting their suitability for CD diagnostic algorithms, at least in Yucatan. Specificity was consistently high across the RDTs, ranging from 93.0% to 100% (95% CIs: 80.9-100). These findings underscore the need to locally validate diagnostic tools. The BiolineTM Chagas Ab and Chagas Sero K-SeT tests met the WHO benchmarks for RDT performance and, if validated in larger cohorts, could strengthen the screening of T. cruzi infection and epidemiological surveillance in Mexico.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Diagnostic Accuracy of the TR Chagas Bio-Manguinhos Rapid Test in the Municipality of Canguçu, Rio Grande do Sul, a Historically Endemic Area for Chagas Disease in Southern Brazil: a prospective study.Acta tropica · 2026
- evaluation of the performance of rapid diagnostic tests in point-of-care conditions for serological diagnosis of chagas disease in pregnant women in Colombia.PLoS neglected tropical diseases · 2026
- From delay to action: rethinking Chagas disease diagnosis with rapid testing in Colombia.Journal of global health · 2026
- Stage-Informed Trypanosoma cruzi Antigen Selection Enhances Assay Concordance in the Evaluation of Discordant Chagas Disease Serology.Microorganisms · 2026
- Using Bayesian methods to evaluate the diagnostic performance of blood-based diagnostic tests for American cutaneous leishmaniasis and Chagas disease in rural Panamanian dogs in the absence of a gold standardVeterinary Parasitology · 2026
- Comparison of Cobas Plasma Separation Cards and Whatman filter paper cards versus conventional serum samples for anti-Trypansoma cruzi antibody detection.PLoS neglected tropical diseases · 2026