Stress-informed parametric shape adaptive patient-specific implant design in computer-assisted jaw reconstruction.
Ankit Nayak, Yu-Xiong Su, Jane Jingya Pu
School of Advanced Engineering, University of Petroleum and Energy Studies (UPES), Dehradun, India. · Division of Oral and Maxillofacial Surgery, Faculty of Dentistry, The University of Hong Kong, Hong Kong, Hong Kong SAR, China.
Key points
From the paper's conclusion- Hence, if the design could analyze these parameters and adjust them accordingly, it could help avoid plate fracture.
Abstract
INTRODUCTION: Reconstructing the jaw after tumor removal or trauma increasingly relies on custom 3D-printed titanium plates. This customization applies to the implant's shape. It usually does not apply to the internal structure. The width and thickness of the plate are still determined by the designer, based on experience and clinical observation. The critical design parameters of the PSI govern its mechanical performance. Hence, if the design could analyze these parameters and adjust them accordingly, it could help avoid plate fracture. This study introduces a parametric-design approach that closes this gap. METHODS: The proposed algorithm addresses these challenges by using stress data derived from finite element analysis (FEA) to selectively reinforce only the high-stress regions of a reconstruction plate, rather than uniformly reinforcing the entire plate. The reinforcement is smoothly integrated into the plate's geometry to prevent abrupt changes in cross-section that could create stress concentrations. The approach was applied to a baseline plate at two regions of elevated stress. RESULTS AND DISCUSSION: Plate A reinforced only the primary zone; this reduced peak stress by 36.0% and strain by 43.8% in that zone, the largest reduction observed across both design variants, and also lowered stress and strain in the second, unmodified zone by 12.7% and 14.3%, respectively. Plate B reinforced both zones; this further reduced stress and strain at the second zone by approximately 30.6% and 33.3%, respectively, and reduced strain at the primary zone by approximately 40.6% relative to baseline. Relative to Plate A, Plate B showed a slightly higher stress at the first zone; this stress remained well below the original baseline value. This work shows that patient-specific implant geometry can be shaped directly by biomechanical feedback. This algorithm offers an efficient computational workflow, enabling implants that are not only anatomically tailored but also mechanically optimized as per given loading conditions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- In silico biomechanical evaluation of personalized A-shaped titanium plates for mandibular condylar base fractures.Computer methods in biomechanics and biomedical engineering · 2026
- Finite Element Analysis of Conventional Fixation and 3D ‐Printed Scaffold Integration for Treating Large Osseous Femoral DefectsJournal of Biomedical Materials Research Part B Applied Biomaterials · 2025
- Finite element analysis of patient specific three-dimensional titanium plate versus conventional two miniplate fixation for mandibular angle fractures.BMC oral health · 2026
- Biofiligree®—A Concept for Bone Fracture Fixation PlatesSci · 2026
- The "PAMFOSS" framework: guidance for the design and application of 3D-printed implants for segmental bone defects.3D printing in medicine · 2026
- Towards minimally invasive prosthetic surgery of the temporomandibular joint: design and integrated CAD-FEM preclinical approach on the mini-pig.Computer methods in biomechanics and biomedical engineering · 2026