A new ridge filter design method for PBS-IMPT
Hongguo Sheng, Guoliang Zhang
Key points
From the paper's conclusion- Background Adopting the new ridge filter (RF) design method enables the use of a single proton beam energy to achieve plan quality comparable to IMPT while reducing treatment time.
Abstract
Background Adopting the new ridge filter (RF) design method enables the use of a single proton beam energy to achieve plan quality comparable to IMPT while reducing treatment time. Purpose The study enhances the RF design, significantly improving its dose modulation capability and further demonstrates the ability to complete more complex plan. Methods A PTV-based optimization method (Monte Carlo calculation + IPOPT solver) was used to first create a standard IMPT plan as reference. RF hexagon was then assembled based on the optimized spot weights according to the new design method. The dose distribution through the RF was compared with that of the reference plan and previous design to validate the new design. The study also evaluated the modulation capability of RF in a simultaneous integrated boost (SIB) plan and analyzed relevant FLASH parameters (DADR, total treatment duration). Results For both the standard IMPT plan and the simultaneous boost plan, the dose distribution through RF hexagon was highly comparable to the reference, with gamma passing rates exceeding 97%. Compared to RF square, the RF hexagon shows a closer similarity to the reference. The entire PTV and ~85% of the oar volume received DADRs higher than 40 Gy/s when the beam intensity reached 1.0 × 10 10 protons/s. When the beam intensity surpassed 5×10 11 protons/s, the total treatment time was reduced to less than 2.66 second. Conclusion Compare to square RF, the new RF design improve its dose modulation capability at single spot. Furthermore, the RF’s performance in the simultaneous boost plan confirms that the new design exhibits modulation capability for more complex dose distributions (versus single-target dose plan).
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Dynamic Range Modulation via Integrated Universal Range Shifter and Ridge Filter: Enabling Simplified Delivery for FLASH-Enabled IMPT.Physics in medicine and biology · 2026
- Ultrafast proton delivery with pin ridge filters (pRFs): repurposing single-energy delivery for motion management in proton therapy.Physics in medicine and biology · 2026
- Single‐field‐uniform‐dose‐per‐fraction simultaneous dose and dose rate optimization (SFUDPF–SDDRO) method for proton FLASH therapyMedical Physics · 2026
- Integrating a Ripple Filter into Pencil Beam Scanning Proton Therapy for Breast Cancer: from Dosimetric Feasibility to Acute Toxicity Assessment.BMC cancer · 2026
- Dynamic Proton arc therapy for paediatric parameningeal rhabdomyosarcomaJournal of Applied Clinical Medical Physics · 2026
- Combining proton FLASH and spatially fractionated radiotherapy: experimental and simulation-based dosimetric characterizationPhysics in Medicine and Biology · 2026