Physics in medicine and biology

Ultrafast proton delivery with pin ridge filters (pRFs): repurposing single-energy delivery for motion management in proton therapy.

Ahmal Jawad Zafar, Xiaofeng Yang, Sunil William Dutta, Yinan Wang, Duncan Henry Bohannon, Zachary Diamond, Pretesh Patel, Sibo Tian,

Department of Radiation Oncology and Winship Cancer Institute, Emory University, Emory Proton therapy center, Atlanta, Georgia, 30308, United States. · Department of Radiology Oncology, Emory University, Clifton Rd, Atlanta, Georgia, 30322-1007, United States.

Key points

From the paper's conclusion
  • This proposed method significantly reduces delivery time and organ-at-risk dose. Future work should focus on planning quality assurance (QA) and clinical implementation in patients with limited breath-hold capacity.

Abstract

OBJECTIVE: Highly conformal proton stereotactic body radiation therapy (SBRT) remains vulnerable to respiratory interplay, and standard mitigation techniques are either patient-dependent or enlarge the irradiated volume. Long delivery times dominated by energy-layer switching (ELS) limit the use of breath-hold (BH). We propose using pin-ridge filters (pRF) to eliminate ELS, shorten beam-on time, and reduce margins. APPROACH: A monoenergetic pRF plan was constructed, with the highest clinical energy (Emax) obtained from the reference internal target volume (ITV)-based intensity modulated proton therapy (IMPT) plan. To avoid per-patient commissioning, we used a bank of pre-commissioned monoenergetic beams spanning 110-200 MeV. Each beam model enables the development of a gross tumor volume (GTV)-based downstream IMPT (IMPT-DS) plan (reduced margins in the BH setting). Next, a nested pencil-beam-direction (PBD) spot-reduction process iteratively removed low-weighted spots from each PBD and, finally, generated pRFs with coarser resolution using a single-energy. This method was validated on five liver and one lung SBRT cases (10 Gy ×5) by comparing the relative total volume receiving the prescription dose (conformity index CI(GTV+5mm) =VRx/VGTV+5mm) and the mean lung-GTV and liver-GTV doses between the conventional IMPT and pRF plans. MAIN RESULTS: Across five liver SBRT cases, the average CI(GTV+5mm) improves from 2.64 (IMPT) to 1.16 (IMPT-pRF). The reduced dose cloud results in ~26% reduction in Dmean to liver-GTV. The mean beam-on time per field dropped from 43.02 s (IMPT) to 7.14 s (IMPT-pRF), i.e., an average saving of 35.88 s per beam (83.4% shorter). Lung SBRT follows a similar trend for CI(GTV+5mm) (3.22 to 1.58), lung-GTV Dmean (10.3 to 6.9 Gy), and average beam-on time (decreased from 47.2 to 9.5s per beam). These dosimetric improvements are primarily due to change in the planning approach (free-breathing ITV to BH GTV), whereas the key contribution of the pRF is shortening of beam-on time, which makes BH delivery feasible and thereby allows this margin reduction. SIGNIFICANCE: This proposed method significantly reduces delivery time and organ-at-risk dose. Future work should focus on planning quality assurance (QA) and clinical implementation in patients with limited breath-hold capacity.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Ultrafast proton delivery with pin ridge filters (pRFs): repurposing single-energy delivery for motion management in proton therapy — Science Skim