Dual-target VIM-PSA deep brain stimulation using octopolar electrodes: surgical technique and feasibility study.
Lucía Triguero-Cueva, José Pablo Martínez-Barbero, Carlos Javier Madrid-Navarro, María José Pérez-Navarro, Adolfo Mínguez-Castellanos, Bartolomé Marín-Romero, Benjamín Iáñez-Velasco, Francisco Escamilla-Sevilla,
Department of Neurology, Hospital Universitario Virgen de las Nieves, Granada, Spain. · Department of Neuroradiology, Hospital Universitario Virgen de las Nieves, Granada, Spain.
Key points
From the paper's conclusion- Single-trajectory dual-target VIM-PSA DBS using octopolar electrodes was successfully performed with acceptable stereotactic accuracy and without major surgery-related adverse events in this small cohort. The approach may increase anatomical coverage and postoperative programming options. Clinical data remain to be determined.
Abstract
BACKGROUND/OBJECTIVES: Deep brain stimulation (DBS) is an established surgical treatment for essential tremor (ET). Both the thalamic ventral intermediate nucleus (VIM) and the posterior subthalamic area (PSA) are effective tremor targets can be approached through a single trajectory. This technical feasibility report describes a single-trajectory dual-target VIM-PSA DBS surgical protocol using octopolar electrodes and evaluates stereotactic accuracy, anatomical contact distribution, and preliminary surgical safety. METHODS: Eleven patients with medication-refractory ET underwent bilateral DBS implantation within the framework of an ongoing randomized, double-blind, crossover clinical trial comparing VIM-DBS and PSA-DBS. Preoperative planning was performed using Brainlab Elements, and postoperative reconstruction was performed using Guide™ XT after fusion of preoperative MRI with one-month postoperative CT. Clinical and electrical data (efficacy outcomes, stimulation thresholds, side-effect thresholds, active contact selection, energy consumption, habituation, patient-reported outcomes, and programming burden) were not assessed in this report. RESULTS: A total of 22 octopolar DBS leads were implanted. Postoperative reconstruction confirmed anatomical contact distribution across the reconstructed VIM model and the PSA/cZI target region. Mean planned-to-implanted deviation across contacts was 1.28 mm for the left electrodes and 1.50 mm for the right electrodes. One patient developed transient symptomatic perielectrode edema; no intracerebral hemorrhage or any other major surgery-related adverse event was observed during early postoperative follow-up. CONCLUSION: Single-trajectory dual-target VIM-PSA DBS using octopolar electrodes was successfully performed with acceptable stereotactic accuracy and without major surgery-related adverse events in this small cohort. The approach may increase anatomical coverage and postoperative programming options. Clinical data remain to be determined.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Deep brain stimulation for arm tremor: one-year blinded follow-up of a randomized trial.Neurological research and practice · 2026
- The Modern Utility of Awake Deep Brain Stimulation SurgeryOperative Neurosurgery · 2026
- Deep Brain Stimulation of the Posterior Subthalamic Area and the Subthalamic Nucleus in Tremor‐Dominant Parkinson's Disease: A Randomized, Crossover TrialMovement Disorders · 2026
- Intraoperative asymmetry of thalamic stimulation mapping in Essential Tremor: specificity of first hemisphere implantation and transitory lesioning effect.Frontiers in neurology · 2026
- Clinical Importance and Usefulness of Direct Targeting of the Ventral Intermediate Nucleus of the Thalamus Using Fluid and White Matter Suppression Imaging at 3T MRI.Operative neurosurgery (Hagerstown, Md.) · 2026