Case report: complete pilot balloon detachment in a ventilator-dependent patient with high-risk tracheostomy exchange.
Xinyu Lai, Yan Huang, Mingyu Hu, Lin Huang
Intensive Care Unit, Xiangya Changde Hospital, Changde, Hunan, China.
Key points
From the paper's conclusion- Conclusion This case highlights recognition of cuff-system failure, specialist escalation, shared risk assessment, and surveillance when exchange is hazardous.
Abstract
BACKGROUND: Complete detachment of a pilot balloon and one-way valve eliminates the standard route for tracheostomy-cuff inflation and pressure measurement. Definitive tube exchange may be hazardous in selected patients. CASE DESCRIPTION: A 74-year-old ventilator-dependent woman with a long-term tracheostomy developed an audible upper-airway leak, a ventilator leak alarm, and reduced displayed tidal volume. Inspection showed complete loss of the pilot balloon and one-way valve, while the residual inflation line remained attached to the tube. Computed tomography documented post-tracheostomy changes but did not report peri-tracheostomy granulation. Otolaryngology bedside examination identified granulation tissue below the stoma and considered tube removal and exchange difficult, with risks of bleeding, aspiration, respiratory difficulty, and loss of the airway. After multidisciplinary assessment and informed shared decision-making, the existing tube was retained and an exceptional bridge manoeuvre was used to access the residual inflation-line lumen, inject approximately 3 mL of air, and occlude the line. Because the pilot balloon and one-way valve were absent, the standard connection for a cuff manometer no longer existed; intracuff pressure was not measured. The reported volume was case-specific, was not a target, and did not establish a safe cuff pressure, efficacy, or safety. Routine hourly observations near the event were not immediate pre-/post-intervention measurements and cannot demonstrate an intervention-related physiological effect. Intermittent leakage persisted, and the defective tube remained in situ at family-requested discharge. CONCLUSION: This case highlights recognition of cuff-system failure, specialist escalation, shared risk assessment, and surveillance when exchange is hazardous. This case-specific manoeuvre should not be interpreted as a recommended bedside procedure. Definitive management remains replacement of the defective tracheostomy tube whenever safely feasible.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Phantom system for endotracheal tube cuff seal testing during mechanical ventilationMedical & Biological Engineering & Computing · 2026
- Pilot study comparing the quality of airway seal in an endotracheal tube with a viscoelastic cuff compared to standard endotracheal tubes.Multidisciplinary respiratory medicine · 2026
- Cuff-induced focal tracheomegaly managed with a custom silicone Y-Stent: A case report.Respiratory medicine case reports · 2026
- Tracheal Laceration and Endotracheal Tube Cuff Rupture During Thyroidectomy: A Case Report and Literature Review.Cureus · 2026
- Sudden Respiratory Impairment due to Simultaneous Airway Obstruction and Tension Pneumothorax in a Patient Who Underwent Tracheostomy.Clinical case reports · 2026
- Progressive Extension of a Post-intubation Tracheal Laceration Presenting As Massive Pneumomediastinum: A Case Report.Cureus · 2026