Phantom system for endotracheal tube cuff seal testing during mechanical ventilation
Delu Zheng, Tamaralayefa B. Agbiki, Ricardo Correia, Cong He, Sergiy Korposh, Sandor Erdody, Barrie Hayes‐Gill, David W. Hewson,
University of Nottingham · Applied Photonics (United Kingdom)
Key points
From the paper's conclusion- As the rigid cylinder specified in EN ISO 5361:2023 does not adequately represent in vivo airway conditions, a benchtop cuff seal testing system was developed incorporating a tracheal model, lung model, mechanical ventilation, and automatic cuff pressure control.
Abstract
Leakage of orogastric secretions past endotracheal tube (ETT) cuffs contributes to ventilator-associated pneumonia (VAP), yet in vivo tracheal seal testing is impractical, requiring phantom evaluation. As the rigid cylinder specified in EN ISO 5361:2023 does not adequately represent in vivo airway conditions, a benchtop cuff seal testing system was developed incorporating a tracheal model, lung model, mechanical ventilation, and automatic cuff pressure control. A rigid cylinder and a bio-inspired tracheal model with tissue-like mechanical properties were evaluated using five high-volume low-pressure (HVLP) polyvinyl chloride (PVC) cuffed ETTs and five polyurethane (PU) cuffed ETTs, with and without mechanical ventilation, and with and without automatic cuff inflation. For PVC cuffed ETTs, fluid leakage rate was highest in the cylinder phantom compared with the bio-inspired phantom, ~ 30% lower without mechanical ventilation and ~ 50% with mechanical ventilation (both p < 0.001). In contrast, PU ETTs exhibited minimal leakage. Long-term testing revealed intracuff pressure decay over 48 h at 37 °C for both PVC (12.24 ± 1.13 cmH2O) and PU (9.06 ± 1.00 cmH2O) cuffed ETTs, while automatic cuff pressure control further reduced leakage for PVC ETTs. The bio-inspired phantom revealed sealing behaviours not captured by cylinder testing, supporting clinically relevant evaluation of ETT sealing performance and cuff materials.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Pilot study comparing the quality of airway seal in an endotracheal tube with a viscoelastic cuff compared to standard endotracheal tubes.Multidisciplinary respiratory medicine · 2026
- The endotracheal cuff sealing–perfusion margin depends on mean arterial pressure: an analytical framework with a contact-mechanics test of its premiseFrontiers in Medicine · 2026
- Influence of different airway devices on intra-arrest ventilation during bag-valve-device ventilation - a prospective randomized controlled cadaver studyCritical Care · 2025
- Polymeric Sealing Materials in Pneumatic Actuators: From Degradation Mechanisms to Failure Diagnostics and Service-Life Prediction.Materials (Basel, Switzerland) · 2026
- Development of a Controlled Pneumatically Actuated Lung Phantom for Surgical Simulation Support and Collapse ModelingCurrent Directions in Biomedical Engineering · 2026
- Personalized automatic management of tracheal cuff pressure and subglottic secretions drainage to prevent pneumonia in critically ill intubated patients. The MICROINHALO multicenter randomized controlled trialIntensive Care Medicine · 2026