Airway stenosis alters the balance between tidal volume, air entrainment, and oxygen-source delivery during normal frequency and superimposed jet ventilation: a bench model study.

Mingyuan Yang, Zhuomin Deng, Hong Li, Jing Guo, Yuxue Yao, Shuwang Yang, Qinghao Cheng

Department of Anesthesiology, Emergency General Hospital, Beijing, China. · PMLS Upstream Marketing Department, Shenzhen Mindray Bio-medical Electronics Co., Ltd, Shenzhen, China.

Key points

From the paper's conclusion
  • Distal FiO 2 did not establish preserved ventilation or arterial oxygenation, and CO 2 elimination was not measured.

Abstract

BACKGROUND: Severe airway stenosis alters gas mixing and volume delivery during open-system jet ventilation. How high-frequency (HF) superimposition and component titration modify these effects remains incompletely characterized. OBJECTIVE: To compare stenosis-induced changes in distal oxygen concentration (FiO2), tidal volume (VT), pressure, flow, and estimated gas composition among pure normal-frequency jet ventilation (NFJV) and superimposed HF jet ventilation (SHFJV) with titration of either the NF (SHFJV-NF) or HF component (SHFJV-HF). METHODS: An open rigid-bronchoscopy bench model was tested with and without 90% fixed airway stenosis. Pure NFJV and SHFJV-NF shared 56 NF parameter settings; SHFJV-HF included 60 HF settings with a fixed NF component. Linear mixed-effects models evaluated strategy × airway interactions and Holm-adjusted pairwise contrasts. A separate matched model assessed HF addition. Gas-component fractions and volumes were estimated at each setting using a two-source mixing model. RESULTS: Strategy × airway interactions were significant for all five measured outcomes (all P < 0.001). Median stenosis-induced changes were +4.7, +29.1, and +24.7 percentage points for FiO2 and -454, -577, and -232 mL for VT during pure NFJV, SHFJV-NF, and SHFJV-HF, respectively. At identical NF settings, HF addition modified the stenosis response by +24.1 percentage points for FiO2, -111.7 mL for VT (95% CI -173.6 to -49.8), +2.02 cmH2O for measured positive end-expiratory pressure (PEEP), and -8.11 L·min-1 for expiratory flow magnitude (all P < 0.001). Estimated entrained-air volume decreased in all strategies, whereas oxygen-source volume did not increase uniformly. CONCLUSION: At identical NF settings, HF addition amplified the stenosis-associated FiO2 rise, VT loss, and PEEP increase. HF-component titration produced a different response across its tested parameter range. Estimated gas-composition changes were consistent with reduced oxygen dilution. Distal FiO2 did not establish preserved ventilation or arterial oxygenation, and CO2 elimination was not measured.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Airway stenosis alters the balance between tidal volume, air entrainment, and oxygen-source delivery during normal frequency and superimposed jet ventilation: a bench model study — Science Skim