Temperature dependence and consistency of dark current measurement methods for infrared focal plane arrays

Rongjie Feng, Jiaping Long, Xinjie Li, Yongyong Cao, Zilong Zeng, Zhenyi Deng, Zixin Liang, Yu Liu,

Key points

From the paper's conclusion
  • These results indicate improved relative BCS–TC consistency over 120 K–150 K while emphasizing that the applicable measurement range is detector and configuration dependent.

Abstract

Dark current is a critical parameter for infrared focal plane detectors, while practical measurement methods may exhibit temperature-dependent discrepancies. This study systematically evaluates the consistency between the blind cold shield (BCS) and theoretical coefficient (TC) methods using paired measurements of 24 mid-wave infrared detector chips at 120 K and 150 K, with 140 K measurements included as an intermediate experimental condition for descriptive analysis. Relative deviation and Bland–Altman analyses were employed to quantify inter-method consistency. At 120 K, 70.8% of chips showed TC dark-current values more than 20% below the BCS baseline, compared with 29.2% at 150 K. The mean absolute relative deviation decreased from 74.9% to 42.1%, while the geometric mean TC/BCS ratio increased from approximately 0.40–0.92. Bland–Altman analysis showed persistent absolute chip-to-chip variability, while the geometric mean TC/BCS ratio moved closer to unity at 150 K. These results indicate improved relative BCS–TC consistency over 120 K–150 K while emphasizing that the applicable measurement range is detector and configuration dependent.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers