Temperature dependence and consistency of dark current measurement methods for infrared focal plane arrays
Rongjie Feng, Jiaping Long, Xinjie Li, Yongyong Cao, Zilong Zeng, Zhenyi Deng, Zixin Liang, Yu Liu,
Key points
From the paper's conclusion- These results indicate improved relative BCS–TC consistency over 120 K–150 K while emphasizing that the applicable measurement range is detector and configuration dependent.
Abstract
Dark current is a critical parameter for infrared focal plane detectors, while practical measurement methods may exhibit temperature-dependent discrepancies. This study systematically evaluates the consistency between the blind cold shield (BCS) and theoretical coefficient (TC) methods using paired measurements of 24 mid-wave infrared detector chips at 120 K and 150 K, with 140 K measurements included as an intermediate experimental condition for descriptive analysis. Relative deviation and Bland–Altman analyses were employed to quantify inter-method consistency. At 120 K, 70.8% of chips showed TC dark-current values more than 20% below the BCS baseline, compared with 29.2% at 150 K. The mean absolute relative deviation decreased from 74.9% to 42.1%, while the geometric mean TC/BCS ratio increased from approximately 0.40–0.92. Bland–Altman analysis showed persistent absolute chip-to-chip variability, while the geometric mean TC/BCS ratio moved closer to unity at 150 K. These results indicate improved relative BCS–TC consistency over 120 K–150 K while emphasizing that the applicable measurement range is detector and configuration dependent.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Mid-wavelength infrared high-operating temperature detector for integrated photonicsApplied Physics Letters · 2026
- Thermographic Sensing for Biomedical Electronic Circuits: Sensor Selection, Acquisition Requirements, and Emissivity Compensation.Sensors (Basel, Switzerland) · 2026
- Linear-mode photon counting performance and intrinsic dark current in HgCdTe APDs for quantum optical applicationsarXiv · 2026
- Photometric parameter inversion and nonlinearity characterization of infrared focal plane arrays based on flat-field data.Optics express · 2026
- Stiles-Crawford effect in infrared two-photon visionBiomedical Optics Express · 2025
- Temperature- and Pressure-Dependent Calibration Transfer Bias of WMS, FWMS, and DAS Methane Retrievals near 1654 nm.Sensors (Basel, Switzerland) · 2026