Temperature- and Pressure-Dependent Calibration Transfer Bias of WMS, FWMS, and DAS Methane Retrievals near 1654 nm.
Murat Guder, Ibrahim T Ozdur
Department of Electrical and Electronics Engineering, TOBB University of Economics and Technology, 43 Sogutozu Caddesi, Ankara 06560, Turkey.
Key points
From the paper's conclusion- Across all 35 operating points, the RMSE is 10.5% for FWMS, 13.4% for WMS, and 14.9% for DAS.
Abstract
Temperature and pressure changes can introduce substantial bias when a tunable diode laser absorption spectroscopy calibration is used away from its reference condition. Here, scanned wavelength modulation spectroscopy (WMS), fixed-wavelength MS (FWMS), and integrated area direct absorption spectroscopy (DAS) are compared for the CH4 R(3) manifold near 1654 nm. To isolate the behavior of the retrieval methods, all three simulations use the same HITRAN2020 line-by-line model, optical depth, laser model, target geometry, and calibration state. The analysis covers temperature variation at constant pressure, pressure variation at constant temperature, and coupled temperature-pressure variation in a closed volume. With temperature held constant, the FWMS bias varies from -5.2% to -0.8% over 0.7-1.1 atm. At constant pressure, it changes from +23.4% to -9.8% between -20 and +40 °C. DAS gives the smallest endpoint bias in the closed-volume case, ranging from +19.1% to -8.0%. Across all 35 operating points, the RMSE is 10.5% for FWMS, 13.4% for WMS, and 14.9% for DAS. The pooled ranking is unchanged for modulation indices between 1.6 and 3.0 and for DAS integration windows between ±10 and ±100 GHz.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Simplified Calibration and Laser Nonlinearity Error Analysis for Peak-Locking TDLAS SystemsPhotonics · 2026
- Influence of key operational parameters on harmonic signals in WMS-TDLAS: a simulation study.Applied optics · 2026
- Baseline-Free Wide-Range Methane Concentration Measurement Based on Logarithmic Second Derivative Spectroscopy.Analytical chemistry · 2026
- Metrological Validation of Low-Cost DS18B20 Digital Temperature Sensors Using the TH-001 Procedure: Calibration Models, Uncertainty, and ReproducibilityMetrology · 2026
- Longitudinally resolved temperature measurement in diode-pumped alkali lasers via dual-wavelength CH4 absorption spectroscopy with one-step fitting.Spectrochimica acta. Part A, Molecular and biomolecular spectroscopy · 2026
- A reconfigurable and compact circular multi-pass sensor for ultra-sensitive methane detectionJournal of Instrumentation · 2026