Effect of distal cantilever and anterior implant diameter on framework and cortical bone stress in a PEEK-based All-on-Four mandibular prosthesis
Shobha Rodrigues, S. Mukherjee, Reon David Sequeira, Krishna Kumar P, Nithesh Naik, Prashant Bajantri, Sowmya N, Shivani Shetty,
Manipal Academy of Higher Education
Key points
From the paper's conclusion- Conclusion: Distal cantilever extension was the primary determinant of biomechanical amplification in a PEEK-based mandibular All-on-Four prosthesis, while reduced anterior implant diameter exerted a secondary but additive effect.
Abstract
Background: Polyetheretherketone (PEEK) has been proposed as an alternative framework material for full-arch implant-supported prostheses due to its lower elastic modulus relative to metallic frameworks. However, quantitative data regarding the combined biomechanical influence of distal cantilever extension and anterior implant diameter in PEEK-based mandibular All-on-Four rehabilitations remain limited. Purpose: To quantify the effects of distal cantilever presence and anterior implant diameter (3.5 mm vs. 4.5 mm) on framework stress and peri-implant cortical bone response in a PEEK-based mandibular All-on-Four prosthesis using three-dimensional finite element analysis. Methods: Four implant configurations were modeled: Standard-NC (4.5 mm anterior, no cantilever), Mixed-NC (3.5/4.5 mm anterior, no cantilever), Standard-C (4.5 mm anterior, with cantilever), and Mixed-C (3.5/4.5 mm anterior, with cantilever). A 300 N static load was applied at premolar and molar positions under vertical (0°) and 30° oblique loading. Peak von Mises stress in the PEEK framework and crestal cortical bone, along with maximum principal cortical strain, were recorded. Results: Distal cantilever extension increased peak framework stress by up to 59% under premolar vertical loading (222.24 MPa vs. 139.60 MPa) and increased crestal cortical stress by approximately 49% (18.36 MPa vs. 12.31 MPa) compared with the non-cantilevered baseline configuration. Under molar vertical loading, maximum principal strain increased by approximately 105% in the Mixed-C model compared to the Standard-NC configuration (0.0080 vs. 0.0039). Reduction of anterior implant diameter increased cortical strain by approximately 28-31% in non-cantilevered models and further amplified strain when combined with cantilever extension. Peak framework stresses reached 222.24 MPa in localized regions. Conclusion: Distal cantilever extension was the primary determinant of biomechanical amplification in a PEEK-based mandibular All-on-Four prosthesis, while reduced anterior implant diameter exerted a secondary but additive effect. These findings highlight the mechanical sensitivity of low-modulus framework systems to cantilever geometry and implant diameter in full-arch rehabilitations.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effect of Connection Geometry and Cantilever Length on the Mechanical Behavior of Ti6Al4V and PEEK Bar-Supported Implant Prostheses with Angled Abutments.Journal of functional biomaterials · 2026
- Clinical and Finite Element Evaluation of Fixed Implant Prosthesis Designs for the Edentulous MandibleClinical Implant Dentistry and Related Research · 2026
- Biomechanical impact of anterior cantilever length on subperiosteal implants with different materials and configurations.Journal of prosthodontics : official journal of the American College of Prosthodontists · 2026
- Mechanical behavior of complete arch implant-supported prostheses with distal cantilevers fabricated from different digital material combinationsJournal of Prosthetic Dentistry · 2026
- PEEK plates for the fixation of mandibular condylar base fractures: a finite element analysis comparison of single and double plate configurations with titanium and resorbable systems.Journal of cranio-maxillo-facial surgery : official publication of the European Association for Cranio-Maxillo-Facial Surgery · 2026
- Biomechanical assessment of titanium, zirconia, and polyetheretherketone abutments under functional loads: A finite element study.Journal of prosthodontic research · 2026