Validation of Azure Kinect for upper limb kinematic assessment after arthroscopic rotator cuff repair: task-dependent accuracy and postoperative overestimation.
Yu-Fei Chen, Chuang-Yu Zhan, Ling-Jie Zhang, Yue-Xin Zhang, Yun-Ru Ma
Department of the First School of Clinical Medicine, The First Affiliated Hospital of Fujian Medical University, Fuzhou, China. · Department of Rehabilitation Medicine, School of Health, Fujian Medical University, Fuzhou, China.
Key points
From the paper's conclusion- During functional tasks, Azure Kinect demonstrated good agreement with IMUs for sagittal-plane movements (ICC2,5 = 0.708 - 0.907 for shoulder flexion; 0.804 - 0.846 for elbow flexion) with good to excellent waveform consistency (CMC > 0.75; CCC > 0.70), but substantially lower agreement and waveform similarity in coronal and horizontal planes. Bland-Altman analysis showed relatively small sagittal-plane errors (SEM < 15°) but larger biases in non-sagittal planes, with proportional bias detected for most joint planes and tasks (p < 0.05). For 3D shoulder ROM assessment, good agreement was observed for flexion (ICC2,5 = 0.837), abduction (ICC2,5 = 0.729), and rotation (ICC2,5 = 0.735). No significant preoperative inter-system differences were found. Postoperatively, Azure Kinect significantly overestimated flexion (137.9 ± 13.5° vs. 97.1 ± 25.1°, p = 0.002) and external rotation (48.6 ± 14.6° vs. 31.0 ± 14.4°, p = 0.038). Both systems detected significant flexion changes pre-to-post, but neither detected changes in abduction or rotation.
Abstract
INTRODUCTION: Arthroscopic rotator cuff repair is the standard treatment for rotator cuff tears (RCTs), yet postoperative functional evaluation continues to rely heavily on subjective clinical scales, with limited objective kinematic assessment. This study evaluated the concurrent validity of an Azure Kinect-based markerless system against a reference inertial measurement unit (IMU) system for upper limb kinematic assessment in patients undergoing RCT repair, and explored its postoperative clinical utility. METHODS: Twelve patients with unilateral full-thickness RCTs were assessed preoperatively and at four months postoperatively. Participants performed six functional activities of daily living tasks and 3D shoulder range-of-motion (ROM) tests, with kinematic data synchronously collected by both systems. Inter-system agreement was evaluated using intraclass correlation coefficients (ICC2,5), Pearson's r, Lin's concordance correlation coefficient (CCC), Bland-Altman analysis, waveform similarity metrics (coefficient of multiple correlation - CMC), and error metrics (root mean square error-RMSE, mean absolute error-MAE, standard error of measurement-SEM). RESULTS: During functional tasks, Azure Kinect demonstrated good agreement with IMUs for sagittal-plane movements (ICC2,5 = 0.708 - 0.907 for shoulder flexion; 0.804 - 0.846 for elbow flexion) with good to excellent waveform consistency (CMC > 0.75; CCC > 0.70), but substantially lower agreement and waveform similarity in coronal and horizontal planes. Bland-Altman analysis showed relatively small sagittal-plane errors (SEM < 15°) but larger biases in non-sagittal planes, with proportional bias detected for most joint planes and tasks (p < 0.05). For 3D shoulder ROM assessment, good agreement was observed for flexion (ICC2,5 = 0.837), abduction (ICC2,5 = 0.729), and rotation (ICC2,5 = 0.735). No significant preoperative inter-system differences were found. Postoperatively, Azure Kinect significantly overestimated flexion (137.9 ± 13.5° vs. 97.1 ± 25.1°, p = 0.002) and external rotation (48.6 ± 14.6° vs. 31.0 ± 14.4°, p = 0.038). Both systems detected significant flexion changes pre-to-post, but neither detected changes in abduction or rotation. DISCUSSION: These preliminary findings indicate task- and plane-dependent validity for the Azure Kinect, with good sagittal performance but substantial errors in multi-planar motions and postoperative overestimation. The system may serve as a screening tool for tracking postoperative functional recovery, but its current accuracy is insufficient for standalone surgical outcome assessment. Results should be interpreted with caution given the small sample size and confirmed in larger cohorts.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of hip and knee range of motion measurements using Azure Kinect and a digital goniometer in elite female weightliftersScientific Reports · 2026
- Concurrent Validity and Agreement of Viewpoint-Dependent Markerless Motion Tracking in Upper-Limb Assessment: A Validation Against Optoelectronic Motion Capture.Sensors (Basel, Switzerland) · 2026
- Validation of View-Specific Markerless Motion Analysis for Lower-Extremity Kinematics After Anterior Cruciate Ligament ReconstructionSensors · 2026
- Measurement Uncertainty and Traceability in Upper Limb Rehabilitation Robotics: A Metrology-Oriented ReviewJournal of Sensor and Actuator Networks · 2026
- Markerless 3D motion capture in return to sport decision making for ACL injured high level athletes: a narrative review with proposed clinical implementation pathway.Annals of joint · 2026
- Markerless On-Device Detection of Compensatory Movement Patterns in Upper-Limb Rehabilitation Exercises from Monocular RGB Video: A Validation Study in Healthy Adults.Sensors (Basel, Switzerland) · 2026