Experimental benchmarking of Geant4 electromagnetic physics lists for X-ray transmission and relative air-kerma profiles in 3D-Printed ABS and PETG mammography phantoms.

Greiciane J Cesário, Crislane J Cesário, Edilaine H Silva, Linda V E Caldas, Carmen C Bueno

Instituto de Pesquisas Energéticas e Nucleares, Comissão Nacional de Energia Nuclear IPEN/CNEN, São Paulo, SP, 05508-000, Brazil; University of São Paulo (USP), Institute of Physics (IF), 05508-090, São Paulo, Brazil. Electronic address: jcesario@usp.br. · Instituto de Pesquisas Energéticas e Nucleares, Comissão Nacional de Energia Nuclear IPEN/CNEN, São Paulo, SP, 05508-000, Brazil.

Key points

From the paper's conclusion
  • This study evaluated four Geant4 electromagnetic physics lists for reproducing X-ray transmission, relative air-kerma profiles, and spectral changes in homogeneous acrylonitrile butadiene styrene (ABS) and glycol-modified polyethylene terephthalate (PETG) phantoms.

Abstract

Three-dimensional (3D) printing is increasingly used to manufacture dosimetric phantoms in diagnostic radiology, requiring computational models that accurately reproduce radiation attenuation. This study evaluated four Geant4 electromagnetic physics lists for reproducing X-ray transmission, relative air-kerma profiles, and spectral changes in homogeneous acrylonitrile butadiene styrene (ABS) and glycol-modified polyethylene terephthalate (PETG) phantoms. The phantoms were manufactured by fused deposition modeling and irradiated with the RQA-M4 mammography beam quality at 35 kVp. Measurements at five depths (0-4 cm) were performed using CaSO4:Dy thermoluminescent dosimeters. The experimental setup was modeled in Geant4 11.3.2 using 108 primary photons and the G4EmLivermorePhysics, G4EmPenelopePhysics, G4EmStandardPhysics, and G4EmStandardPhysics_option4 lists. Air kerma was calculated from photon fluence crossing virtual scoring surfaces at the measurement depths. All physics lists reproduced the decrease in transmission with depth and the greater attenuation in PETG. Mean absolute percentage differences ranged from 3.6% to 4.2% for ABS and from 7.3% to 8.5% for PETG. G4EmLivermorePhysics provided the smallest mean difference for ABS, whereas G4EmStandardPhysics provided the smallest for PETG. The largest local discrepancies were 12.6% for ABS at 4 cm and 14.0% for PETG at 2 cm. The spectra showed fluence reduction, beam hardening, and strong overlap among the physics lists. Displacements of ±0.5 mm produced maximum sensitivities of 2.4% for ABS and 4.1% for PETG, below the largest simulation measurement discrepancies. Geant4 reproduced the overall attenuation and spectral-modification trends, although quantitative agreement depended on material and depth.