Concurrent validity of a novel digital dynamometerfor assessing performance on the Cranio-Cervical Flexion Test.
Marco Barbero, Daniele Lonati, Marco Griseri, Corrado Cescon, Alessandro Bonafine, Tiziano Negrisolo, Emanuele Falzone, Samuel Pedrucci,
Rehabilitation Research Laboratory 2rLab, Department of Business Economics, Health and Social Care, University of Applied Sciences and Arts of Southern Switzerland, Manno - Switzerland. · MedAction, Rapperswil-Jona - Switzerland.
Key points
From the paper's conclusion- These findings demonstrate strong concurrent validity between the NOD and the PBU, supporting the NOD as a valid digital alternative for assessment of the CCFT.
Abstract
INTRODUCTION: The Cranio-Cervical Flexion Test (CCFT) is a validated clinical test used to evaluate deep cervical flexor muscle function. A pressure biofeedback unit (PBU) has traditionally been the reference device. Recently, a new portable digital device known as the Neuromuscular Cranio-cervical Device (NOD) has been developed as a potential alternative. METHODS: In this cross-sectional concurrent validation study, 88 participants were recruited from four outpatient physiotherapy practices, including individuals with current neck pain and individuals with a previous history of neck pain. All participants performed the CCFT using both devices. The PBU was inflated to a baseline pressure of 20 mmHg, with the participant then performing C-CF to increase the pressure in increments of 2 mmHg over five progressive stages, while the NOD was calibrated to corresponding force targets. Agreement and concurrent validity between the devices were assessed, and C-CF range of motion (ROM) was analyzed using linear mixed-effects models. RESULTS: Agreement between CCFT performance scores obtained with the NOD and the PBU was 94.3%, with a very strong positive correlation (Spearman's ρ = 0.930, p < 0.001). The NOD measured slightly greater C-CF ROM than the PBU (mean difference 0.68°), with significant differences at lower CCFT stages that converged at higher stages of the test. CONCLUSION: These findings demonstrate strong concurrent validity between the NOD and the PBU, supporting the NOD as a valid digital alternative for assessment of the CCFT.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Standardized Protocol Combining McKenzie Mechanical Therapy and Neurodynamic Techniques for Cervical Radiculopathy.Journal of visualized experiments : JoVE · 2026
- The Validity and Reliability of a Novel Testing Device and Six‐Way Neck‐Strength Testing Protocol for Peak Force and Rate of Force DevelopmentEuropean Journal of Sport Science · 2026
- A Portable Force Measurement Device for Lip and Tongue Muscles.Journal of oral rehabilitation · 2026
- Thermal Evaluation Protocol for Vocal Function: Development, Content Validity, and Reliability.Journal of voice : official journal of the Voice Foundation · 2026
- Reliability and Validity of the Single Breath Count and Breath Holding Time Dysfunctional Breathing Variables in Patients with Non-Specific Chronic Neck Pain.Medical sciences (Basel, Switzerland) · 2026
- Construct validity of real-world digital mobility outcomes in patients after proximal femoral fracture: a cross-sectional observational studyScientific Reports · 2026