A Portable Force Measurement Device for Lip and Tongue Muscles.
Yutong Guo, Weining Wang, Bo Rong Xing, Hu Qiao
Key Laboratory of Shaanxi Province for Craniofacial Precision Medicine Research, College of Stomatology, Xi'an Jiaotong University, Xi'an, China. · Independent Researcher in Artificial Intelligence, Xi'an, China.
Key points
From the paper's conclusion- The portable device demonstrated good laboratory calibration performance and high short-term repeatability. Preliminary known-groups discrimination was observed between healthy children and children with orofacial muscle dysfunction. Further studies using established reference methods, inter-operator assessment and larger samples are required before criterion validity, diagnostic performance and routine clinical applicability can be established.
Abstract
BACKGROUND: Quantitative assessment of lip and tongue muscle function remains limited, as current approaches mainly rely on pressure-based instruments or clinical observation. OBJECTIVE (S): To develop a portable device for independently measuring upper lip, lower lip and tongue force and to evaluate its laboratory calibration performance, short-term repeatability and preliminary known-groups discrimination in children. METHODS: Standard weights ranging from 100 to 5000 g were used to assess device linearity and laboratory repeatability. Short-term repeatability was evaluated in 18 healthy children using three measurements performed at 1-h intervals. A preliminary known-groups comparison was conducted between 18 healthy children and 18 children clinically classified as having orofacial muscle dysfunction. Explosive force, peak force and endurance were measured at the tongue tip, mid-tongue, upper lip and lower lip. RESULTS: Laboratory calibration demonstrated coefficients of determination of 0.999, absolute residuals within 0.1 N and coefficients of variation ranging from 0.00% to 0.591%. Intraclass correlation coefficients for the 12 measured parameters ranged from 0.856 to 0.992, indicating high short-term repeatability. Children with clinically diagnosed orofacial muscle dysfunction showed lower explosive force, peak force and endurance than healthy controls at all four measurement sites. All 12 between-group comparisons remained statistically significant after Bonferroni correction (p < 0.0042). CONCLUSION: The portable device demonstrated good laboratory calibration performance and high short-term repeatability. Preliminary known-groups discrimination was observed between healthy children and children with orofacial muscle dysfunction. Further studies using established reference methods, inter-operator assessment and larger samples are required before criterion validity, diagnostic performance and routine clinical applicability can be established.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Orofacial Strength in a Typically Developing Pediatric Population: A Comparison of the Iowa Oral Performance Instrument and Tongueometer.Journal of speech, language, and hearing research : JSLHR · 2026
- Evaluation of Bite Force Measurements Using Sensor-Based Systems: A Feasibility Study.Journal of oral rehabilitation · 2026
- Pilot validation of a piezo‑resistive gnathodynamometer for maximum voluntary bite force measurement in adults: a cross sectional comparative study.BMC oral health · 2026
- Concurrent validity of a novel digital dynamometerfor assessing performance on the Cranio-Cervical Flexion Test.Archives of physiotherapy · 2026
- Digital handgrip assessment in musculoskeletal health: reliability, measurement precision, interchangeability, and clinical implications for rehabilitation and sports medicineBMC Sports Science, Medicine and Rehabilitation · 2026
- Risk Classification Instrument for Orofacial Myofunctional Disorders - ICR/DMO: Reliability (Part III).Journal of applied oral science : revista FOB · 2026