Sim&Size versus IFU-Based Sizing for the Artisse 2.0 Intrasaccular Device in a Multicenter Registry.
Fernanda Rodriguez-Erazú, Ricardo A Hanel, Otavio F de Toledo, Salvador F Gutierrez-Aguirre, Jonathan Cortese, Markus A Möhlenbruch, Saleh Lamin, Riitta Rautio,
From the Lyerly Neurosurgery (F.R.-E., R.A.H., O.F.D.T., S.F.G.-A.), Baptist Neurological Institute, Jacksonville, Florida, USA; Jacksonville University (F.R.-E., O.F.D.T., S.F.G.-A.), Research Department 4121 Jacksonville University Jacksonville, FL, US 32211-3321; Department of Interventional Neuroradiology (NEURI vascular center) (F.R.-E., J.C., L.S.), Bicêtre University-Hospital, Le Kremlin-Bicetre, France; Paris-Saclay University Faculty of Medicine (J.C., L.S.), INSERM U1195, Le Kremlin-Bicetre, France; Department of Neuroradiology (M.A.M.), Heidelberg University Hospital, Heidelberg, Germany; Interventional Neuroradiology. Radiology (S.L.), Queen Elizabeth Hospital, Birmingham, UK; Department of Interventional Radiology (R.R.), Turku University Hospital and Turku University, Turku, Finland; Department of Neuroradiology (J.F.), University Medical Center Hamburg-Eppendorf, Hamburg, Germany and Department of Neurology/Institute of Neurointervention (M.K.-O.), Paracelsus Medical University, Salzburg, Austria. · From the Lyerly Neurosurgery (F.R.-E., R.A.H., O.F.D.T., S.F.G.-A.), Baptist Neurological Institute, Jacksonville, Florida, USA; Jacksonville University (F.R.-E., O.F.D.T., S.F.G.-A.), Research Department 4121 Jacksonville University Jacksonville, FL, US 32211-3321; Department of Interventional Neuroradiology (NEURI vascular center) (F.R.-E., J.C., L.S.), Bicêtre University-Hospital, Le Kremlin-Bicetre, France; Paris-Saclay University Faculty of Medicine (J.C., L.S.), INSERM U1195, Le Kremlin-Bicetre, France; Department of Neuroradiology (M.A.M.), Heidelberg University Hospital, Heidelberg, Germany; Interventional Neuroradiology. Radiology (S.L.), Queen Elizabeth Hospital, Birmingham, UK; Department of Interventional Radiology (R.R.), Turku University Hospital and Turku University, Turku, Finland; Department of Neuroradiology (J.F.), University Medical Center Hamburg-Eppendorf, Hamburg, Germany and Department of Neurology/Institute of Neurointervention (M.K.-O.), Paracelsus Medical University, Salzburg, Austria. drricardohanel@icloud.com.
Key points
From the paper's conclusion- For the Artisse 2.0, Sim&Size and chart-based sizing selected the same device in most cases. Sim&Size use was associated with higher technical success, whereas first-device success and procedure and fluoroscopy times were comparable. Both approaches achieved high procedural success with a low complication rate and may be used according to operator preference, with simulation potentially most useful in anatomies at the limits of the device range.
Abstract
BACKGROUND AND PURPOSE: Virtual simulation is increasingly promoted for intrasaccular device sizing, but whether it changes device selection or procedural outcomes for new, highly conformable devices is unknown. We compared Sim&Size simulation with instructions-for-use (IFU) chart sizing for the Artisse 2.0 in early multicenter experience. MATERIALS AND METHODS: We retrospectively analyzed data from the INSPIRE-A registry (November 2022-June 2024). Sim&Size use depended on software availability. We assessed agreement between sizing tools, technical success, first-device success, adjunctive devices, procedure and fluoroscopy times, complications, and the cause of each technical failure. RESULTS: Sim&Size was used in 88/123 cases (71.5%). When both methods were used (n=59), the same device was selected in 54 (91.5%). Technical success was 98.9% with Sim&Size and 85.7% without (p=0.007), whereas first-device success (86.4% vs 80.0%, p=0.41) and procedure and fluoroscopy times did not differ. Perioperative complications occurred in 4/123 (3.3%). Of the six technical failures, five were related to device fit, mostly in anatomies at the limits of the device range, and one resulted from a perforation before deployment. A dome-to-neck ratio <1 was associated with failure (3/7 vs 3/116, p=0.002). CONCLUSIONS: For the Artisse 2.0, Sim&Size and chart-based sizing selected the same device in most cases. Sim&Size use was associated with higher technical success, whereas first-device success and procedure and fluoroscopy times were comparable. Both approaches achieved high procedural success with a low complication rate and may be used according to operator preference, with simulation potentially most useful in anatomies at the limits of the device range.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Feasibility, Technical Safety, and Early Clinical Experience with an Automated Annular Suturing Device in Minimally Invasive Valve Replacement.Journal of clinical medicine · 2026
- Simulating Vein Compressibility for Ultrasound-Guided Femoral Access: Speed, Accuracy, and Phantom ValidationarXiv · 2026
- Comparison of the Technical Success of Obtura and StarClose Systems for Femoral Arterial Access-Site Closure in Neurointerventional Procedures: A Single-Center Retrospective Cohort Study.Journal of clinical medicine · 2026
- From precision to strength: computer vision for suture quality assessment—an ex vivo pilot studySurgical Endoscopy · 2025
- Stitch or seal? Suture-based versus bioabsorbable plug closure devices in transfemoral stroke thrombectomy: a multicenter cohort study.Journal of neurointerventional surgery · 2026
- Accuracy of Metallic-Sleeved Versus Sleeve-Free Surgical Guides in Static Computer-Assisted Implant Surgery: A Systematic Review and Meta-Analysis.Journal of clinical medicine · 2026