Sim&Size versus IFU-Based Sizing for the Artisse 2.0 Intrasaccular Device in a Multicenter Registry.

Fernanda Rodriguez-Erazú, Ricardo A Hanel, Otavio F de Toledo, Salvador F Gutierrez-Aguirre, Jonathan Cortese, Markus A Möhlenbruch, Saleh Lamin, Riitta Rautio,

From the Lyerly Neurosurgery (F.R.-E., R.A.H., O.F.D.T., S.F.G.-A.), Baptist Neurological Institute, Jacksonville, Florida, USA; Jacksonville University (F.R.-E., O.F.D.T., S.F.G.-A.), Research Department 4121 Jacksonville University Jacksonville, FL, US 32211-3321; Department of Interventional Neuroradiology (NEURI vascular center) (F.R.-E., J.C., L.S.), Bicêtre University-Hospital, Le Kremlin-Bicetre, France; Paris-Saclay University Faculty of Medicine (J.C., L.S.), INSERM U1195, Le Kremlin-Bicetre, France; Department of Neuroradiology (M.A.M.), Heidelberg University Hospital, Heidelberg, Germany; Interventional Neuroradiology. Radiology (S.L.), Queen Elizabeth Hospital, Birmingham, UK; Department of Interventional Radiology (R.R.), Turku University Hospital and Turku University, Turku, Finland; Department of Neuroradiology (J.F.), University Medical Center Hamburg-Eppendorf, Hamburg, Germany and Department of Neurology/Institute of Neurointervention (M.K.-O.), Paracelsus Medical University, Salzburg, Austria. · From the Lyerly Neurosurgery (F.R.-E., R.A.H., O.F.D.T., S.F.G.-A.), Baptist Neurological Institute, Jacksonville, Florida, USA; Jacksonville University (F.R.-E., O.F.D.T., S.F.G.-A.), Research Department 4121 Jacksonville University Jacksonville, FL, US 32211-3321; Department of Interventional Neuroradiology (NEURI vascular center) (F.R.-E., J.C., L.S.), Bicêtre University-Hospital, Le Kremlin-Bicetre, France; Paris-Saclay University Faculty of Medicine (J.C., L.S.), INSERM U1195, Le Kremlin-Bicetre, France; Department of Neuroradiology (M.A.M.), Heidelberg University Hospital, Heidelberg, Germany; Interventional Neuroradiology. Radiology (S.L.), Queen Elizabeth Hospital, Birmingham, UK; Department of Interventional Radiology (R.R.), Turku University Hospital and Turku University, Turku, Finland; Department of Neuroradiology (J.F.), University Medical Center Hamburg-Eppendorf, Hamburg, Germany and Department of Neurology/Institute of Neurointervention (M.K.-O.), Paracelsus Medical University, Salzburg, Austria. drricardohanel@icloud.com.

Observational

Key points

From the paper's conclusion
  • For the Artisse 2.0, Sim&Size and chart-based sizing selected the same device in most cases. Sim&Size use was associated with higher technical success, whereas first-device success and procedure and fluoroscopy times were comparable. Both approaches achieved high procedural success with a low complication rate and may be used according to operator preference, with simulation potentially most useful in anatomies at the limits of the device range.

Abstract

BACKGROUND AND PURPOSE: Virtual simulation is increasingly promoted for intrasaccular device sizing, but whether it changes device selection or procedural outcomes for new, highly conformable devices is unknown. We compared Sim&Size simulation with instructions-for-use (IFU) chart sizing for the Artisse 2.0 in early multicenter experience. MATERIALS AND METHODS: We retrospectively analyzed data from the INSPIRE-A registry (November 2022-June 2024). Sim&Size use depended on software availability. We assessed agreement between sizing tools, technical success, first-device success, adjunctive devices, procedure and fluoroscopy times, complications, and the cause of each technical failure. RESULTS: Sim&Size was used in 88/123 cases (71.5%). When both methods were used (n=59), the same device was selected in 54 (91.5%). Technical success was 98.9% with Sim&Size and 85.7% without (p=0.007), whereas first-device success (86.4% vs 80.0%, p=0.41) and procedure and fluoroscopy times did not differ. Perioperative complications occurred in 4/123 (3.3%). Of the six technical failures, five were related to device fit, mostly in anatomies at the limits of the device range, and one resulted from a perforation before deployment. A dome-to-neck ratio <1 was associated with failure (3/7 vs 3/116, p=0.002). CONCLUSIONS: For the Artisse 2.0, Sim&Size and chart-based sizing selected the same device in most cases. Sim&Size use was associated with higher technical success, whereas first-device success and procedure and fluoroscopy times were comparable. Both approaches achieved high procedural success with a low complication rate and may be used according to operator preference, with simulation potentially most useful in anatomies at the limits of the device range.

Read the paper