AJNR. American journal of neuroradiology

Deep Learning-Accelerated 3D T2 CUBE MRI of the Cochlea at 1.5T and 3T: Sonic DL With AIR Recon DL Versus AIR Recon DL Alone.

Katherine L Reinshagen, Stacy Hooker, Nisha Nayak, Dominic Juliano, Eugene Milshteyn, Ajeetkumar Gaddipati, Mary Thomas, Dan Rettmann,

Ear Harvard Medical School (K.L.R., S.H., I.J.), Harvard Medical School, GE HealthCare (E.M., M.T.), Boston, MA, USA; GE HealthCare (A.G.), Waukesha, WI, USA; GE HealthCare (D.R.), Rochester, MN, USA; Wentworth-Douglass Hospital (I.J.), Enterprise Service Group - Radiology, Mass General Brigham, Dover, NH, USA and Department of Radiology (I.J.), University of Turku, Turku, Finland. · Ear Harvard Medical School (K.L.R., S.H., I.J.), Harvard Medical School, GE HealthCare (E.M., M.T.), Boston, MA, USA; GE HealthCare (A.G.), Waukesha, WI, USA; GE HealthCare (D.R.), Rochester, MN, USA; Wentworth-Douglass Hospital (I.J.), Enterprise Service Group - Radiology, Mass General Brigham, Dover, NH, USA and Department of Radiology (I.J.), University of Turku, Turku, Finland. ijambor@mgb.org.

Computational

Key points

From the paper's conclusion
  • A combined Sonic DL acceleration with ARDL-denoised reconstruction for 3D T2 CUBE produced significantly higher subjective image quality at both 1.5T and 3T, and, at 3T, higher inter-reader reliability than standard CUBE-ARDL, while shortening acquisition time by 34%-38%. These findings support further evaluation, including diagnostic accuracy studies, of CUBE-SDL as an alternative to CUBE-ARDL.

Abstract

BACKGROUND AND PURPOSE: Three-dimensional fast spin-echo MR cisternography (CUBE) of the internal auditory canal (IAC) and cerebellopontine angle (CPA) requires longer acquisition times that can limit clinical throughput. We evaluated whether a deep learning (DL)-based highly accelerated reconstruction technique, Sonic DL`™ (SDL), combined with a commercially available DL denoising reconstruction (AIR™ Recon DL, ARDL), preserves or improves subjective image quality and cochlear anatomy visualization relative to standard CUBE-ARDL while shortening scan time. MATERIALS AND METHODS: Consecutive patients underwent brain MRI for suspected IAC pathology at 1.5T and 3T with paired acquisition of standard 3D CUBE-ARDL and an accelerated 3D CUBE-SDL sequence. Both used the same voxel size at 1.5T (0.6x0.6x1.2 mm3) and 3T (0.6x0.6x0.6 mm3). Scan times for CUBE-ARDL versus CUBE-SDL were 2:33 versus 1:41 min at 1.5T and 1:55 versus 1:11 min at 3T. Two independent neuroradiologists blinded to MRI acquisition method scored randomized, de-identified images on a 5-point Likert scale across five domains (overall image quality, noise, sharpness, artifacts, cochlear anatomy visualization). Reader-averaged scores for CUBE-ARDL and CUBE-SDL were compared with paired Wilcoxon signed-rank tests, with reader-specific and cumulative link mixed model (CLMM) sensitivity analyses and false discovery rate (FDR) adjustment; intraclass correlation coefficients (ICC) quantified inter-reader reliability for each technique; Gwet's AC2 and weighted (Conger's) kappa with linear ordinal weights quantified pooled inter-reader agreement. RESULTS: After exclusion of 5 training cases per field strength, 49 (1.5T) and 34 (3T) patients were analyzed. CUBE-SDL scored significantly better than CUBE-ARDL for all five domains at both field strengths (all FDR-adjusted P < .001), consistently for each reader and in the CLMM. Inter-reader reliability (ICC) for CUBE-SDL was moderate-to-excellent at 3T (0.63-0.92) and moderate at 1.5T (0.65-0.74), generally exceeding CUBE-ARDL (3T: 0.35-0.61; 1.5T: 0.50-0.71). Pooled inter-reader agreement (Gwet's AC2) ranged from 0.56 to 0.89. CONCLUSIONS: A combined Sonic DL acceleration with ARDL-denoised reconstruction for 3D T2 CUBE produced significantly higher subjective image quality at both 1.5T and 3T, and, at 3T, higher inter-reader reliability than standard CUBE-ARDL, while shortening acquisition time by 34%-38%. These findings support further evaluation, including diagnostic accuracy studies, of CUBE-SDL as an alternative to CUBE-ARDL.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers