Differential pulse voltammetric assay of H 1 -antihistamine fexofenadine in pharmaceutical formulations using a bare glassy carbon electrode

Monika Čerňanská, Romana Králiková, Jana Blaškovičová, Ľubomír Švorc

Key points

From the paper's conclusion
  • A simple differential pulse voltammetric procedure was developed for the determination of the H 1 -antihistamine fexofenadine in commercial pharmaceutical formulations using a bare glassy carbon electrode.

Abstract

A simple differential pulse voltammetric procedure was developed for the determination of the H 1 -antihistamine fexofenadine in commercial pharmaceutical formulations using a bare glassy carbon electrode. The electrochemical behavior of fexofenadine was investigated by cyclic voltammetry in phosphate buffer solutions. The analyte provided an irreversible anodic signal at approximately +0.75 V versus an Ag/AgCl/3 mol·L −1 KCl reference electrode in phosphate buffer at pH 10, where the highest current response was obtained. The influence of scan rate indicated that the oxidation process was predominantly diffusion-controlled. Differential pulse voltammetry was optimized and then used for quantitative analysis. Under the optimized conditions, the method provided a linear concentration range from 9.5 to 67.9 μmol·L −1 , a detection limit of 2.7 μmol·L −1 , and repeatability expressed as a relative standard deviation of 4.8 % for twelve replicate measurements. The influence of selected tablet excipients, including glucose, sucrose, starch, and cellulose, was evaluated, with changes in the integrated fexofenadine response remaining below 5 % at all tested mass ratios. The proposed procedure was applied to the assay of two commercial fexofenadine tablet products, Allegra ® 120 mg and Ewofex ® 120 mg, using the standard addition method, providing recoveries of 99.4 % and 98.7 %, respectively. The method offers a simple option for the differential pulse voltammetric assay of fexofenadine in tablet formulations.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers