Differential Pulse Voltammetric Analysis of Cannabidiol on Fluorine-Doped Tin Oxide (FTO) Electrode Interface: Quantification in Pharmaceutical Oil with Assessment of Electrode Surface Stability.
Mario Romero, Eliana Rocha Fontalvo, María L Ospina-Castro, Victoria A Arana, Carlos Mario Meléndez, Jolián Andrés Vargas Álzate, Andrea Ramos-Hernández
Grupo de Investigación Química Supramolecular Aplicada, Programa de Química, Facultad de Ciencias Básicas, Universidad del Atlántico, Carrera 30 No. 8-49, Puerto Colombia 081007, Colombia. · Grupo de Investigación Ciencias, Educación y Tecnología-CETIC, Programa de Química, Facultad de Ciencias Básicas, Universidad del Atlántico, Carrera 30 No. 8-49, Puerto Colombia 081007, Colombia.
Key points
From the paper's conclusion- The FTO electrode exhibited high reproducibility, withstanding more than 20 consecutive additions to the same electrode without loss of response, thanks to its intrinsic stability and the strategic use of forced convection.
Abstract
This work presents the use of an unmodified fluorine-doped tin oxide (FTO) electrode as an electrochemical sensor for the stable and highly reproducible detection of cannabidiol (CBD) in an oral pharmaceutical formulation (CBD-OS). This was achieved using an electrochemical methodology with 0.1 M LiClO4 in acetonitrile as the electrolyte medium. The oxidation process of CBD on the FTO was extensively studied using cyclic voltammetry to analyze the influence of forced convection during the experiment. Quantification was performed by differential pulse voltammetry with standard addition to the dissolved CBD oral oil (CBD-OS), which allowed for the determination of a detection limit of 4.8 µM and a correlation coefficient R2 = 0.9989. Statistical tests indicated no significant differences between the proposed electrochemical methodology and HPLC. The FTO electrode exhibited high reproducibility, withstanding more than 20 consecutive additions to the same electrode without loss of response, thanks to its intrinsic stability and the strategic use of forced convection. The use of an unmodified electrode enabled the development of a simple, reproducible, robust, and cost-effective analytical methodology for CBD analysis in pharmaceutical samples.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Theophylline electrochemical sensing in pharmaceutical drugs using disposable boron-doped diamond thin film electrodesFunctional Diamond · 2025
- Portable voltammetric method for the simultaneous determination of Δ9-Tetrahydrocannabinol and cannabidiol in alkaline medium using unmodified screen-printed carbon electrodesTalanta · 2026
- “On–off-on” fluorescence sensing of ascorbic acid based on a Fe3+-modulated carbon quantum dots–rhodamine 6G dual-emission system: application to pharmaceutical quality controlScientific Reports · 2026
- Rapid carbamazepine detection by means of a disposable electrochemical sensor based on a molecularly imprinted polymerTalanta · 2025
- Validated QuEChERS-HPLC-UV Method for Quantification of Cannabidiol in Pharmaceutical Nanoformulations.ChemPlusChem · 2026
- Electrochemical sensing of sildenafil in pharmaceutical formulations using an (S,N)-doped reduced graphene oxide-modified electrode.Beilstein journal of nanotechnology · 2026