Portable blood glucose meters do not reliably detect hypoglycemia in ferrets (Mustela putorius furo) compared with a point-of-care chemistry analyzer.
Kimberly Rollins, Mark Rishniw, Thomas M Donnelly, Logan Kopp
Priest Lake Veterinary Hospital, Nashville, TN. · Veterinary Information Network, Davis, CA.
Key points
From the paper's conclusion- The v-PBGM canine and feline settings agreed more closely with the reference analyzer than did the h-PBGM. However, agreement was insufficient to support diagnosis or management of hypoglycemia based solely on portable glucometer results in ferrets.
Abstract
OBJECTIVE: To compare blood glucose concentrations measured by a veterinary portable blood glucose meter (v-PBGM; AlphaTrak3; Zoetis Inc), a human portable blood glucose meter (h-PBGM; FreeStyle Precision Neo; Abbott Diabetes Care), and a point-of-care chemistry analyzer (reference analyzer; VetScan VS2; Zoetis Inc) in client-owned ferrets. METHODS: Residual heparinized jugular venous blood samples from ferrets undergoing unrelated veterinary evaluations were analyzed from June 2025 through April 2026 within 10 minutes of collection. Blood glucose was measured with the h-PBGM and with the v-PBGM using equine, canine, and feline settings. Agreement was assessed by Bland-Altman analysis and at hypoglycemia thresholds of ≤ 60 and < 70 mg/dL. RESULTS: 24 client-owned domestic ferrets (Mustela putorius furo) were included. Mean blood glucose concentration by the reference analyzer was 90.8 mg/dL (range, 43 to 157 mg/dL). The h-PBGM had the greatest positive fixed bias. The v-PBGM canine and feline settings showed minimal fixed bias, but all glucometers had wide limits of agreement. At the ≤ 60-mg/dL threshold, simple classification agreement was 79% for the canine and feline settings, 75% for the equine setting, and 58% for the h-PBGM. CONCLUSIONS: The v-PBGM canine and feline settings agreed more closely with the reference analyzer than did the h-PBGM. However, agreement was insufficient to support diagnosis or management of hypoglycemia based solely on portable glucometer results in ferrets. CLINICAL RELEVANCE: Low, borderline, unexpected, or treatment-altering blood glucose results should be confirmed with a point-of-care or reference laboratory chemistry analyzer.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of a portable glucometer for blood glucose measurement in juvenile green sea turtles (Chelonia mydas).Australian veterinary journal · 2026
- Agreement of a Handheld Glucose Meter with Laboratory Serum Glucose Measurements in Dairy CowsDairy · 2026
- Comparison of Blood Glucose Measurement Methods at Low Concentrations in Neonates at Risk for Hypoglycemia.The journal of applied laboratory medicine · 2026
- Glucometer Versus Standard Enzymatic Method for Gestational Diabetes Mellitus: A Systematic Review for Diagnostic Test Accuracy.International journal for quality in health care : journal of the International Society for Quality in Health Care · 2026
- Validation of Point-of-care Testing Device HemoCue® Glucose 201RT in Blood Glucose Estimation.Indian journal of community medicine : official publication of Indian Association of Preventive & Social Medicine · 2026
- Hematologic and biochemical prognostic indicators of influenza disease severity in naïve and preimmune ferret (Mustela putorius furo) modelsScientific Reports · 2026