Assessment of Wrist-Worn Activity Tracker Accuracy for Measuring Sedentary Behavior in Outpatients With Cardiovascular Disease: Validation Study.
Koichi Naito, Kazuhiro P Izawa, Noriaki Maeda, Yuya Kasai, Haruna Tani, Misaki Kondo, Hajime Iwama
Department of Physical Therapy, Faculty of Medical Science, Nagoya Aoi University, 3-40 Shioji-cho, Mizuho-ku, Nagoya, 4678610, Japan, 81 52-838-5026. · Department of Health Sciences, Graduate School of Medicine, Kobe University, Kobe, Japan.
Key points
From the paper's conclusion- In older Japanese outpatients with cardiac issues, a consumer-grade wrist-worn tracker overestimated sedentary time by approximately 62 minutes within a standardized 12-hour daytime window compared with a waist-worn comparator, with fixed but no proportional bias. A simple constant correction markedly improved group-level concordance (cross-validated CCC≈0.90), supporting calibrated use of consumer wearables for scalable monitoring in ambulatory cardiac care. However, wide limits of agreement constrain individual-level precision, and the population-specific correction factor requires independent validation before transfer to other settings.
Abstract
BACKGROUND: Wearable monitoring of sedentary behavior is increasingly used in digital cardiovascular care, but the validity of consumer-grade wrist-worn trackers in older patients with cardiac issues is unclear. OBJECTIVE: This study aimed to evaluate the agreement between sedentary time measured by a consumer-grade wrist-worn tracker (Fitbit Inspire 3; Fitbit LLC) and a research-grade waist-worn accelerometer (ActiveStyle Pro HJA-750C; Omron Healthcare Co, Ltd) used as the comparator, in Japanese outpatients with cardiac issues under free-living conditions. METHODS: Fifty-one outpatients with heart failure (28/51, 55%) or ischemic heart disease (23/51, 45%; mean age 78.4, SD 9.0 years; 25/51, 49% male) simultaneously wore both devices for 7 consecutive days. Sedentary time was assessed within a standardized 12-hour daytime window (9 AM to 9 PM) at 1-minute epochs. Agreement was examined using Pearson correlation, Bland-Altman analysis (after confirming normality of the paired differences), Lin concordance correlation coefficient (CCC), and Cohen kappa for distribution-based tertile classification. A constant bias correction was applied, with leave-one-out cross-validation (LOOCV) used as an internal-consistency check on the stability of the sample mean bias. Moderators of the bias were explored using multivariable regression. RESULTS: Mean sedentary time within the 12-hour window was 497.9 (SD 86.5) minutes for the comparator and 559.9 (SD 84.3) minutes for the Fitbit. The devices were strongly correlated (r=0.905, 95% CI 0.838-0.945; P<.001). The Fitbit showed a significant fixed bias of 62.0 minutes (95% CI 51.5-72.5; P<.001) with no proportional bias (slope -0.03; P=.67); the 95% limits of agreement were wide (-11.3 to 135.2 minutes). Uncorrected concordance was moderate (CCC=0.71, 95% CI 0.57-0.80) and improved to high after subtracting the 62-minute bias (CCC=0.904, 95% CI 0.84-0.94); LOOCV indicated that this in-sample improvement was internally stable (out-of-sample CCC=0.90). No candidate factor (sex, age, diagnosis, and musculoskeletal disorder) significantly moderated the bias (model P=.12). Tertile classification showed moderate agreement (κ=0.588, 95% CI 0.405-0.772; 37/51, 72.5% agreement), with all misclassifications between adjacent categories. CONCLUSIONS: In older Japanese outpatients with cardiac issues, a consumer-grade wrist-worn tracker overestimated sedentary time by approximately 62 minutes within a standardized 12-hour daytime window compared with a waist-worn comparator, with fixed but no proportional bias. A simple constant correction markedly improved group-level concordance (cross-validated CCC≈0.90), supporting calibrated use of consumer wearables for scalable monitoring in ambulatory cardiac care. However, wide limits of agreement constrain individual-level precision, and the population-specific correction factor requires independent validation before transfer to other settings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Differential Agreement of Self-Reported Sedentary Behavior and Physical Activity With Accelerometer Measures in Adults With Congenital Heart Disease.Circulation reports · 2026
- Fitbit Sense 2 in hospitalized general medicine patients: A pilot study assessing feasibility and validity of specific measures.International journal of medical informatics · 2026
- Absolute Agreement Between a Consumer Wearable and Research Grade Accelerometry in Free-Living Conditions: Comparison Study.JMIR formative research · 2026
- Accuracy in Step Counting Is Not Accuracy in Quantifying Walking Activity: A Criterion Validity Study of Wearable Devices Under Free-Living Conditions.Medicine and science in sports and exercise · 2026
- Physical Activity and Sedentary Time in Chronic Stroke: An Agreement Study Between the IPAQ-SF and ActiGraph Measures.NeuroRehabilitation · 2026
- A Head-to-Head Comparison of Three Literature Algorithms for Physical Activity Endpoints from a Wrist Accelerometer in Free-Living Conditions.Sensors (Basel, Switzerland) · 2026