Repeatability of Anterior Segment Measurements Using Anterior Segment OCT Combined with Placido Corneal Topography in Dry Eye Patients.
Abdulaziz Alsomali, Musab Abdulaziz Alzoman, Mohammed Falah Aljasir, Abdullah Abdulrahman Alfrayyan, Jumana Talal Hariri, Mohammed Hassan Alam
Ophthalmology Department, College of Medicine, King Faisal University, Alahsa, Saudi Arabia. · Ophthalmology Department, Dhahran Eye Specialist Hospital, Dhahran, Saudi Arabia.
Key points
From the paper's conclusion- The MS-39 demonstrates high intra-session repeatability for most anterior segment parameters in dry eye disease. Corneal astigmatism showed greater variability; therefore, astigmatism readings should be interpreted cautiously in patients with ocular surface instability. This study assessed intra-session repeatability only and did not evaluate inter-session reproducibility or performance relative to healthy controls.
Abstract
PURPOSE: To evaluate the repeatability of anterior segment parameters using the MS-39 in patients with dry eye disease. METHODS: This prospective cross-sectional observational study included 52 eyes from 26 participants with dry eye disease. Each eye underwent three consecutive MS-39 acquisitions within 15 minutes. Repeatability was analyzed using the coefficient of repeatability, coefficient of variation, intraclass correlation coefficient, and repeated-measures analysis of variance. RESULTS: Most parameters showed excellent intra-session repeatability (intraclass correlation >0.98, coefficient of variation <15%), except corneal astigmatism (coefficient of variation approximately 52%). Exploratory stratified analyses suggested lower repeatability in eyes with higher Ocular Surface Disease Index scores and lower Tear Break-Up Time values, particularly for corneal astigmatism. CONCLUSION: The MS-39 demonstrates high intra-session repeatability for most anterior segment parameters in dry eye disease. Corneal astigmatism showed greater variability; therefore, astigmatism readings should be interpreted cautiously in patients with ocular surface instability. This study assessed intra-session repeatability only and did not evaluate inter-session reproducibility or performance relative to healthy controls.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effect of Dry Eye and Ocular Surface Treatment on Agreement and Repeatability of Corneal Astigmatism Measurements Among Three Devices.Journal of cataract and refractive surgery · 2026
- Establishing a precision reference benchmark for corneal epithelial aberration measurements using a new hybrid device across keratoconus stages.Eye and vision (London, England) · 2026
- Repeatability of an SD-OCT biometer (MS-39 AL) and agreement with a validated biometer (IOLMaster 700).Journal of cataract and refractive surgery · 2026
- Repeatability and Reproducibility of the Dry Eye Analyzer in Participants With and Without Symptoms of Dry Eye.Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians (Optometrists) · 2026
- Dynamic corneal responses from successive tonometry at different resting intervals using the Corvis ST in young myopesFrontiers in Bioengineering and Biotechnology · 2026
- The Effect of Dry Eye Treatment on Biomechanics in Refractive Surgery Candidates.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026