Effect of Dry Eye and Ocular Surface Treatment on Agreement and Repeatability of Corneal Astigmatism Measurements Among Three Devices.
Songlian Wang, Xuanqiao Lin, Runhan Shi, Lifang Bai, Dongjin Qian, Jin Yang
University of Shanghai for Science and Technology, Shanghai, People's Republic of China. · Department of Ophthalmology, Eye, Ear, Nose, and Throat Hospital of Fudan University, Shanghai, People's Republic of China.
Key points
From the paper's conclusion- DED was associated with reduced agreement and repeatability of corneal astigmatism measurements obtained using IOLMaster 700, Pentacam, and iTrace. Short-term ocular surface treatment was associated with improved measurement stability.
Abstract
PURPOSE: To evaluate the effect of dry eye disease (DED) and ocular surface treatment on the agreement and repeatability of corneal astigmatism measurements obtained using IOLMaster 700, Pentacam, and iTrace in patients with age-related cataract. SETTING: Eye, Ear, Nose, and Throat Hospital of Fudan University, Shanghai, China. DESIGN: Prospective, single-center study. METHODS: A total of 313 eyes of patients with age-related cataract were included. Eyes with DED received 0.3% sodium hyaluronate eye drops for 2 weeks and were re-examined after treatment. Corneal astigmatism measurements were obtained using IOLMaster 700, Pentacam, and iTrace before pupil dilation. The main parameters included keratometric corneal astigmatism, total corneal astigmatism, and their vector components. Interdevice agreement was assessed using Bland-Altman analysis, and test-retest repeatability was evaluated using within-subject standard deviation, test-retest repeatability, and intraclass correlation coefficients. RESULTS: Compared with the control group, eyes at DED baseline showed wider limits of agreement and poorer TRT for both scalar astigmatism parameters and vector components. After ocular surface treatment, LoA width decreased by 50.5%-62.5% for KCA and 19.9%-29.0% for TCA across the three device comparisons. KCA repeatability improved significantly after treatment for IOLMaster 700 and iTrace, whereas significant improvement in TCA repeatability was observed only for IOLMaster 700. Device- and parameter-specific differences persisted after treatment. CONCLUSIONS: DED was associated with reduced agreement and repeatability of corneal astigmatism measurements obtained using IOLMaster 700, Pentacam, and iTrace. Short-term ocular surface treatment was associated with improved measurement stability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Repeatability of Anterior Segment Measurements Using Anterior Segment OCT Combined with Placido Corneal Topography in Dry Eye Patients.Clinical ophthalmology (Auckland, N.Z.) · 2026
- The impacts of ocular surface factors on refractive outcomes and visual quality in cataract surgery.Current opinion in ophthalmology · 2026
- Modulation of Total Corneal Astigmatism, Corneal Thickness, and Corneal Higher-Order Aberrations Following Tear-Enhancing Drop Instillation: A Retrospective Comparative Study.Ophthalmology and therapy · 2026
- The Effect of Dry Eye Treatment on Biomechanics in Refractive Surgery Candidates.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Establishing a precision reference benchmark for corneal epithelial aberration measurements using a new hybrid device across keratoconus stages.Eye and vision (London, England) · 2026
- Effects of 3% diquafosol sodium on tear film stability, corneal aberration parameters, and vision-related quality of life in pediatric orthokeratology-associated dry eye diseaseBMC Ophthalmology · 2026