Analytical validation of an ultrasensitive multiplexed immunoassay for blood-based biomarkers GFAP, NF-L and tau.
Catherine Demos, Nikhil Padmanabhan, Jermaine Brown, Taron Gorham, Daniel Romero, Rachel Cohen, Jason Abraham, Brian Ngo,
Key points
From the paper's conclusion- Here we demonstrate analytical validation of the MSD® S-PLEX Human Neurology Panel 1, an ultrasensitive multiplex immunoassay for GFAP, NF-L and total tau.
Abstract
Background Blood based biomarkers are increasingly used for non-invasive diagnostics, prognostics, stratification, and therapeutic monitoring, however clinical validation of new biomarkers is often limited by inconsistent assay methods. Assay developers can enable more reliable clinical studies by conducting analytical evaluation to characterize assay performance. In the context of brain injury and neurodegeneration, GFAP, NF-L and tau have emerged as potentially useful biomarkers for stratifying injury severity. Here we demonstrate analytical validation of the MSD® S-PLEX Human Neurology Panel 1, an ultrasensitive multiplex immunoassay for GFAP, NF-L and total tau. Methods Analytical validation testing consisted of multi-site, multi-lot precision, accuracy, stability, dynamic range, and interference. All testing was conducted with plasma, serum, and diluent-based control samples. The precision study was conducted at three sites: Meso Scale Diagnostics, LLC. (Rockville, MD), Johns Hopkins University (Baltimore, MD) and All Children's Hospital (Tampa, FL). Three kit lots were tested at all three sites for a fully crossed study design. All runs contained an 8-point duplicate calibrator curve, duplicate sample measurements, and control samples. Data were analyzed at each site and then transmitted to MSD for combined analysis. Results There were no significant differences between reagent-lot measurements. Intra-plate %CV averaged less than 10%. Between-site comparison of GFAP, NF-L and tau measurements revealed a small but significant difference in measurements across study sites. Accuracy reported that native samples recover within ± 20% of expected concentration through 20-fold dilution, and spiked samples recover within ± 20% through 256-fold dilution. No substances in a panel of potential interferents affected quantitation by more than ± 20% except for erythropoietin. Kits are shelf stable through 24 months, and GFAP, NF-L and tau in samples can withstand at least five freeze/thaw cycles, storage for 24 hours at 4°C, and storage for four hours at ambient temperature. Conclusions The MSD S-PLEX® Human Neurology Panel 1 has been analytically validated for research use biomarker studies measuring GFAP, NF-L and tau. Characterization of sources of variability have identified minor to moderate inter-site variability, but little to no inter-lot variability. Further work is needed to identify an international reference standard and to evaluate use in specific research contexts.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Plasma brain-derived tau: analytical and clinical validation of the first commercial immunoassayScientific Reports · 2026
- A-359 Comparison of Whole Blood and Plasma GFAP and UCH-L1 Blood-Based Biomarkers for Screening Patients with Mild Traumatic Brain InjuryClinical Chemistry · 2026
- Simoa and MSD platforms show analytical discordance but comparable diagnostic performance for GFAP, NF-L, and T-tau in adolescent concussionScientific Reports · 2026
- B-152 Evaluation of the Analytical Performance of Neurofilament Light Chain immunoassaysClinical Chemistry · 2026
- A multivariate plasma biomarker model for glioblastoma exclusion: a retrospective case-control study.Translational oncology · 2026
- B-057 From Plasma to Whole Blood: Validation of GFAP and NFL as Accessible Biomarkers in StrokeClinical Chemistry · 2026